To restore the stored settings to your TIGER Select, first ensure that your instrument is fully booted
and
connected to your PC as described previously
. Ensure that your instrument is in normal
operation, is not in an alarm condition and that no data logging or Health and Safety readings are
being collected.
Important
: Be aware that this process will replace all setup and calibration files.
Click
Restore
against the relevant snapshot.
On the restore pop-up that is then displayed:
Click
Yes
to load the snapshot. When the loading is complete, click
Close
and restart your TIGER
Select. Your TIGER Select will now be restored to the settings and calibration data stored at the time
of that snapshot.
The snapshot menu also allows stored data to be reviewed when an instrument is not attached.
Expand the menu until the relevant snapshot is displayed. Double-click on the snapshot. All the data
stored in that snapshot can now be viewed.