background image

 

TIGER Select

 

Instrument User Manual V3.2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Unrivalled Gas 

Detect

ion.

 

 

ionscience.com

 

 

Содержание Tiger Select

Страница 1: ...TIGER Select Instrument User Manual V3 2 Unrivalled Gas Detection ionscience com...

Страница 2: ...nty Thank you for purchasing your ION Science instrument The standard warranty of your instrument can be extended to five years To receive your extended warranty you must register your instrument onli...

Страница 3: ...TIGER Select Instrument User Manual V3 2 Unrivalled Gas Detection Page 3 of 73 ionscience com Warning...

Страница 4: ...TIGER Select Instrument User Manual V3 2 Unrivalled Gas Detection Page 4 of 73 ionscience com Avertissement...

Страница 5: ...Battery Packs 14 Keypad Function Descriptions 14 3 Getting Started 15 Fitting the Inlet Probe 15 Switching your TIGER Select On 15 Switching your TIGER Select Off 15 Checking the Battery Charge 15 Se...

Страница 6: ...t 32 Downloading Data Logged Readings and Instrument Settings 32 Taking Snapshots 33 Instrument Summary Screen 34 Features Screen 35 Configuration Screen 36 Gas Table Screen 44 Firmware Screen 45 Data...

Страница 7: ...e the Electrode Stack 62 Removing the MiniPID 2 Sensor 63 Lamp Removal and Examination 64 Lamp Cleaning 64 Replacing a Lamp 65 Replacing the MiniPID Electrode Stack 66 9 Troubleshooting 67 Diagnostics...

Страница 8: ...manual We reserve the right at any time and without any notice to remove amend or vary any of the content which appears in this manual Symbols WARNING USED TO INDICATE DANGER WARNINGS WHERE THERE IS A...

Страница 9: ...nents may impair intrinsic safety Do not service the instrument live Remove the Battery Pack before servicing Ingress Protection Continuous exposure to wet weather conditions should be limited to less...

Страница 10: ...TIGER Select Instrument User Manual V3 2 Unrivalled Gas Detection Page 10 of 73 ionscience com Ion Science Ltd offer a take back service Please contact us for more information...

Страница 11: ...ght source many gases associated with benzene are ignored due to their higher eV Throughout the measurement process TIGER Select continues to display real time data ensuring the final reading represen...

Страница 12: ...lled Gas Detection Page 12 of 73 ionscience com 1 Charge Cradle 2 Instrument Boot 3 Alarm LEDs Torch 4 Probe Assembly 5 Display Screen 6 Keypad 7 Instrument Body 8 Battery Pack 9 Charging Contacts 10...

Страница 13: ...ty readings We recommend you install the software and set up your TIGER Select as described in TigerPC Software TIGER Select Kit ION Science provides a range of TIGER Select kits and accessories The c...

Страница 14: ...See Batteries for further details Keypad Function Descriptions The functionality of the A and B keys varies Prompts on the display indicate their functionality at a given time The Up and Down keys ar...

Страница 15: ...display will then show the instrument checking that the Lamp is working the symbol is displayed Next if the instrument is set to Zero when switched on the screen will display the symbol The main runni...

Страница 16: ...chargeable Battery Pack For further details of Battery Packs and recharging refer to Batteries Select the Gas Next ensure that the instrument is set to detect the correct gas This is displayed on the...

Страница 17: ...rticular functions are currently tied to the A and B keys 4 The area between the two soft key indicators displays the measurement units and the gas that the instrument is currently set to detect The T...

Страница 18: ...ro to three sound projection lines are displayed to indicate the volume level there are four sound levels no line is displayed for the quietest Alarm Bell When a Low alarm is triggered the bell symbol...

Страница 19: ...Operational Modes The TIGER Select has three operational modes Standard Running Mode TAC Mode and Tube Mode In Standard Running Mode the display shows a live reading of gas concentration This mode wil...

Страница 20: ...u can set your instrument to operate using the factory calibrated values These are set by ION Science Ltd or authorised service centres Custom You can carry out a Custom Calibration with the 10 0eV Mi...

Страница 21: ...fer to the TigerPC Configuration Screen topic for details of these options Select the preferred option with the Up and Down keys and confirm with Enter For percentage volume press Enter again to confi...

Страница 22: ...updating as the current peak reading changes Press the Reset soft key to reset the peak value Press Esc to cancel Information Pressing the Information soft key provides access to a list of screens wh...

Страница 23: ...s no response factor or is incompatible with the units selected a warning symbol will appear The message can be cleared by pressing the Enter key The gas will still be selected Average Press the Avera...

Страница 24: ...ill be temporarily displayed before TAC Mode is exited Health Safety Only available if the relevant upgrade has been installed Press the Health Safety soft key to display the Health Safety symbol with...

Страница 25: ...setup A tick briefly appears beside the question mark to confirm that Data Logging has started The memory status icon flashes once per second while continuous Data Logging is in progress To stop Data...

Страница 26: ...ter capacity reaches its limit the tube will change colour from bright orange to a green brown If the colour changes beyond the marking the benzene concentration may not be accurate If the colour chan...

Страница 27: ...e cap of the Tube Holder and insert the tube into the Tube Holder Cap ensuring the black arrow on the tube is pointing towards the instrument Screw the Tube Holder Cap and the tube on to the instrumen...

Страница 28: ...the remaining test time and a live STEL reading will be displayed Note that this live reading is for indication only It is only the final reading at the end of the test cycle that should be used as r...

Страница 29: ...ulty please contact your local distributor or contact ION Science directly http ionscience com contact Connecting a TIGER Select to a PC The TIGER Select must be functionally checked prior to entering...

Страница 30: ...tection Page 30 of 73 ionscience com Double click the Tiger icon on your desktop or in the Start menu to open TigerPC or run Tiger exe in the installation folder The Home page will be displayed The me...

Страница 31: ...een This screen will inform you if a new version of PC software is available You can then install the upgrade if required We recommend that TigerPC is kept up to date to ensure that the latest functio...

Страница 32: ...d logged Your instrument will be added to this list once TigerPC has read it as described next Downloading Data Logged Readings and Instrument Settings Click Instrument s The page that is then display...

Страница 33: ...nlogged instrument by TigerPC If not already listed the instrument IRN will be added to the list of instruments Taking Snapshots The first time you read an instrument a snapshot is taken automatically...

Страница 34: ...ection Page 34 of 73 ionscience com Instrument Summary Screen Once an instrument has been read expand its entry in the menu and click on Summary to display the Instrument Summary screen This screen sh...

Страница 35: ...eatures Screen This screen indicates which of the available feature upgrades have been added to your TIGER Select To purchase additional features contact ION Science Ltd or your distributor Once the t...

Страница 36: ...ser Manual V3 2 Unrivalled Gas Detection Page 36 of 73 ionscience com Configuration Screen Use this screen to configure your TIGER Select Click these buttons highlighted above and below to display or...

Страница 37: ...e 37 of 73 ionscience com Note that the images in this topic display the default TigerPC settings Backlight Choose from Permanently off Permanently on On in low ambient light On for limited time Enter...

Страница 38: ...rms continue to sound even if the gas level falls below the threshold until the Esc key is pressed Volume is adjusted using the slider Vibrate Use to set your TIGER Select to vibrate under alarm condi...

Страница 39: ...f operation the TIGER Select starts in o Basic Always start in Basic Start up Mode o Adv Always start in Advanced Start up Mode o Ask Instrument asks the user during start up Calibration Select the ca...

Страница 40: ...save these settings to your TIGER Select The calibration procedure is detailed in Calibration Datalog Select one of the following min and sec set the interval between readings Users can select to reco...

Страница 41: ...he instrument s soft key B When connected to a PC unless it is in TAC Mode the instrument will go into Sleep Mode The instrument will awaken when disconnected from the PC Zeroing The following options...

Страница 42: ...lowing The instrument will alarm when the memory log is full Logging will stop Logging will continue New data will overwrite the oldest data in the log Clock Select 24 or 12 hour format If required se...

Страница 43: ...controls the order in which they are listed in the soft key menus TWA STEL Select the regulatory code to which you are working and specify TWA and STEL periods Instrument ID Enter the text to be disp...

Страница 44: ...ionscience com Gas Table Screen Select Gas Table from the menu to display the Gas Table screen You can modify this table and download it to your instrument To add new gases to the table or delete the...

Страница 45: ...ment select Send to Instrument from the drop down menu shown above Firmware Screen This screen displays the current firmware version installed on your TIGER Select and provides facilities to install n...

Страница 46: ...ta during the process take a snapshot of the instrument before upgrading This can be reloaded after the update is complete Select Send to Instrument to install them on your instrument The following me...

Страница 47: ...ct the TIGER Select at this point Follow the instructions below before disconnecting After the upgrade been sent to the TIGER Select the instrument will install it The TIGER Select will flash the two...

Страница 48: ...instrument is read The Datalog sub menu displays a list of the Datalog sessions Click on a session Details of the data collected during that session are presented in numerical and graphical form on t...

Страница 49: ...e latest Health and Safety readings held on your TIGER Select if the relevant feature upgrade has been installed New readings are downloaded from the TIGER Select when the instrument is read Click Men...

Страница 50: ...your TIGER Select at a point in time If required your TIGER Select can be reset to a snapshot s settings The Snapshots screen lists those stored on the PC Details of a snapshot can be viewed by expan...

Страница 51: ...t Be aware that this process will replace all setup and calibration files Click Restore against the relevant snapshot On the restore pop up that is then displayed Click Yes to load the snapshot When t...

Страница 52: ...TIGER Select Instrument User Manual V3 2 Unrivalled Gas Detection Page 52 of 73 ionscience com Click Delete to remove a selected snapshot...

Страница 53: ...Sensor are brought back to factory specifications and new Factory Calibration data is loaded PID Custom Calibration is performed by the instrument user and should be carried out as part of regular ins...

Страница 54: ...he Calibration Kit are available and ready for use Never calibrate the Zero with the Span gas connected PID Custom Calibration Procedure 1 From the main running screen select the CAL icon on key B 2 H...

Страница 55: ...hat the SPAN 1 has been accepted Press Enter For a two point calibration the calibration procedure is complete 7 SPAN 2 calibration For a three point calibration the gas and concentration for SPAN 2 a...

Страница 56: ...not use abrasive or chemical detergents to clean the instrument as this may reduce the antistatic properties of the materials used clean it using a damp cloth only TIGER Select Parts Diagram 1 Instru...

Страница 57: ...wo types of battery packs are available for use with the TIGER Select instrument a lithium ion Rechargeable Battery Pack and non rechargeable AA Battery Pack to be fitted with 3 AA Alkaline batteries...

Страница 58: ...at if the instrument is left for one year unused it needs to be charged fully before it can continue to be kept in storage Please repeat annually Ensure the TIGER Select is charged for at least 7 hour...

Страница 59: ...e the instrument shuts down Battery fully charged Only use the Charge Cradle supplied with your TIGER Select ION Science Ltd recommends keeping your TIGER Select on charge at all times when not in use...

Страница 60: ...Body and secure it using the clip 5 Replace the Instrument Boot ensuring that it is snug around the front of the Instrument Body and the lights are not obscured 6 Charge the TIGER Select for 7 hours...

Страница 61: ...hts are not obscured WEEE REGULATIONS Dispose of used batteries in accordance with all local and national safety and environmental requirements Inlet Probe Assembly All parts of the probe assembly can...

Страница 62: ...r Internal components must be handled with clean hands and clean tools The Lamp is fragile Handle with great care Never touch the window and do not drop When to Clean or Replace the Lamp The TIGER Sel...

Страница 63: ...clean environment so that the sensor parts will not be contaminated by dust oil or grease On the Sensor Cover loosen the centre screw turning it with a small coin or a suitable flat bladed screwdrive...

Страница 64: ...tamination on the detection window This appears as a blue hue To check for this hold the Lamp in front of a light source and look across the window surface at an angle If necessary clean the Lamp Lamp...

Страница 65: ...conds 4 Remove the residual powder with a short blast of air from the can of clean dry oil free air 5 Refit the Lamp into the Electrode Stack as described previously Replacing a Lamp Never refit a dam...

Страница 66: ...rside of the Electrode Stack The Lamp window should now sit flush within the Electrode Stack This procedure ensures that the Lamp is firmly held in the Electrode Stack with the detection window level...

Страница 67: ...connections are sound and the indicator lights on the charger are in order If the battery fails to charge fit another battery pack if available If using alkaline batteries replace them If the fault pe...

Страница 68: ...o 61010 1 Standards BS EN 60079 0 2012 Electrical apparatus for explosive gas atmospheres General requirements BS EN 60079 11 2012 Explosive atmospheres Equipment protection by intrinsic safety i BS E...

Страница 69: ...sibility on the date this product accompanied by this declaration is placed on the market the product conforms to all technical and regulatory requirements of the above listed directives Name Graham D...

Страница 70: ...indication of benzene breakthrough is displayed in real time Accuracy 10 display reading one digit benzene Linearity 5 display reading one digit Intrinsically Safe Approvals II 1G Ex ia IIC T4 Ga Tamb...

Страница 71: ...06 CFR 47 2008 Class A Weight and Dimensions probe fitted no tube attached Instrument weight 0 75 kg Packed weight 5 kg 176 oz Size 465 mm 18 3 H x 89 mm 3 5 W x 61 mm 2 4 D Warranty Standard Warranty...

Страница 72: ...4 14480 Fax 49 2104 144825 Web www ism d de Email info ism d de Italian Office ION Science Italia Via delle Querce 1 g 40011 Anzola dell Emilia Bologna ITALY Tel 39 051 0561850 Fax 39 051 0561851 Web...

Страница 73: ...ion Updated Contact Information p51 Text changes pp 11 12 18 peak hold 30 help screen 32 features screen 43 electrode stack 49 instrument body 52 Accuracy Linearity 24 08 2018 V0 8 11 V1 0 11 3 2 Upda...