ARA DOCK4 User Manual V1.0
P
ion
eering Gas Sensing Technology.
Page 14 of 15
ionscience.com
Troubleshooting
Problem
Solution
The ARA DOCK4 does not activate.
Connect the power adapter.
Bump test fails and gas leaks out.
Confirm that all gas connections are
attached correctly.
Bump test fails and gas pressure gauge is
almost zero
Replace the empty gas cylinder.
Gas cylinder does not fit to tubing and cannot
be installed to ARA DOCK4.
Inspect the tubing and replace with new one if
it is damaged.
Power Buttons is blinking red
Check if the USB Memory is properly installed.
The audible alarm test fails during bump test or
calibration.
Clean the audible alarm hole
The visual alarm test fails during bump test or
calibration.
1) clean the ARA unit’s LEDs
2) clean the visual alarm hole in test bays