                                                                                                      ARA DOCK4 User Manual V1.0 

 

P

ion

eering Gas Sensing Technology.

 

Page 12 of 15 

ionscience.com

 

Management of the ARA DOCK4 

Replace gas cylinder 

If you need to test gas detectors that is different from the gas type currently installed, or if the gas 
cylinder runs out of gas due to continuous use, you will need to replace the gas cylinder.  

WARNING

:

 

A small amount of gas may leak when gas cylinder is removed. Remove it in a well-

ventilated area.  

WARNING

: Continuous use of the ARA DOCK4 when a gas cylinder is empty may damage the ARA 

DOCK4. Ensure that gas cylinder is replaced when appropriate. 

Connect Communication Cable for Dock4Manager software 

To use the Dock4Manager to manage the ARA DOCK4, a communication cable must be connected. 
The ARA DOCK4 has an ethernet and USB port. Either ports can be used to connect to  a PC. 

Dock4Manager Software Features 

Change gas configurations 

If a calibration gas cylinder of different gases or concentrations is only available which differs from 
the recommended default calibration gases, then these span values need to be changed and sent 
(write) to the ARA DOCK4 via Dock4Manager.  

Install ARA Gas Detector Firmware Upgrade Image 

To perform a firmware upgrade to ARA gas detectors via ARA DOCK4, the firmware upgrade image 
file must be installed in the ARA DOCK4. Users can easily upload the file to ARA DOCK4 by Using 
Dock4Manager. 

Firmware Upgrade of the ARA DOCK4 

Users can update ARA DOCK4 as latest version using Dock4Manager. 

Read Event/Test Logs 

Dock4Manager can read event/test logs from ARA DOCK4, generate reports and forward via email. It 

can also print test and calibration certificates.  

 

There are many other features. Please refer to Dock4Manager Manual. 

 

 

 

Содержание shawcity ARA DOCK4

Страница 1: ...er Manual V1 0 Pioneering Gas Sensing Technology ionscience com Distributed by 91 91 92 Shrivenham Hundred Business Pa Watchfield Oxfordshire SN6 8TY Tel Email Web 01367 899558 info shawcity co uk sha...

Страница 2: ...rging 8 Calibration Gas Installation 8 Set Ventilation Flow Path 9 Operate the ARA DOCK4 10 Turn on off the ARA DOCK4 10 Set gas detectors on Test Bays 11 Start Tests 11 Management of the ARA DOCK4 12...

Страница 3: ...at any time and without any notice to remove amend or vary any of the content which appears in this manual Symbols WARNING USED TO INDICATE DANGER WARNINGS WHERE THERE IS A RISK OF INJURY OR DEATH Ca...

Страница 4: ...exercise due care and follow the appropriate regulations when disposing Always adhere to local regulations and procedures when disposing of the equipment ION Science Ltd offers a take back service Ple...

Страница 5: ...two power sources A 12V DC mains power supply and an internal rechargeable battery The internal rechargeable battery provides approximately 5 hours of continuous operation based on 4 ARA s inserted T...

Страница 6: ...coupler Used to supply fresh air Exhaust Outlet 1 8 Hose Barb coupler Gas and fresh air used in the test are exhausted here Calibration Gas Pressure Gauge Gauge for checking Gas remaining Max 1 500 p...

Страница 7: ...on IrDA communication Sensors Audio and optical Tests Bump Calibration Data Transfer Audible alarm visual alarm Gases Available H2S 25ppm CO 100ppm O2 18 SO2 10ppm Adjustable Gas Cylinder Compatible w...

Страница 8: ...l gas for calibrations The equipment must be attached to a venting system or be used in a well ventilated area Charging The ARA DOCK4 is powered by External 12Vdc Adapter and internal rechargeable bat...

Страница 9: ...RNING Ensure that all calibration gas cylinders are in good condition WARNING Ensure that all calibration gas cylinders contain enough gas WARNING This equipment uses toxic gas for calibrations The eq...

Страница 10: ...ARA DOCK4 If it is powered the Power Button will light up according to the ARA DOCK4status 1 The ARA DOCK4 can be operated when the Power Button lights up green 2 The ARA DOCK4 will automatically turn...

Страница 11: ...alibration Green Calibrates the detectors measurement accuracy against the applied gas 1 After gas detectors are installed press Bump Test Button of Calibration Button to start tests 2 A bump test wil...

Страница 12: ...s an ethernet and USB port Either ports can be used to connect to a PC Dock4Manager Software Features Change gas configurations If a calibration gas cylinder of different gases or concentrations is on...

Страница 13: ...of the equipment 3 Do not attempt to disassemble adjust or service the equipment unless instructions are provided to perform a procedure or a component is listed as a user replaceable component in the...

Страница 14: ...eplace the empty gas cylinder Gas cylinder does not fit to tubing and cannot be installed to ARA DOCK4 Inspect the tubing and replace with new one if it is damaged Power Buttons is blinking red Check...

Страница 15: ...ce com in Email kschari ionscience com ION Science Inc USA Office Tel 1 877 864 7710 Web https ionscience com usa Email info ionscienceusa com ION Science Italy Italy Office Tel 39 051 0561850 Web www...

Отзывы: