                                    SF6 P2 Instrument User Manual V1.2 

 

Unrivalled Gas Detect

ion.

                

Page 5 of 16

                                                      

ionscience.com

 

 

 

 

Wiring 

 
Mains power connects to the orange clamp block, according to the designation printed on the PCB 
near each clamp. 
 
For an optional audible and/or visible alarm device, a voltage free switchover relay output is supplied. 
It is available on the blue 3 position clamp block in the middle. Contacts are, from left to right: N.O. / 
COM / N.C. 
 
During normal operation, COM connects to N.O. During an alarm or error condition, COM disconnects 
from N.O. and switches over to N.C. The same applies in case of power failure, and whenever the 
instrument is taken into setup mode via the host computer interface via the RS232 connector. 
 
The leftmost 6 position clamp block is provided for network operation. Physically, the network follows 
RS485 standards and hence has three wires named A, B, and GND. Their positions are marked on the 
PCB next to the clamps. They must be wired to the corresponding pins on the SF6 NETWORKCONTROLLER P2NC. 

Double clamps are provided for each wire for easy “daisy chaining” of several instrument. 

 

 

 

 
 

 
 
 

 
 
 
 
 
 
 

RS 485 daisy chain PIN B 

RS 485 daisy chain PIN A 

RS 485 daisy chain PIN GND 

Alarm Relay PIN 

“Normally Open” 

Alarm Relay PIN “Common” 

Alarm Relay PIN “Normally Closed” 

RS 232 Connector 

Supply Voltage “EARTH” 

Supply Voltage “L1” 

Supply Voltage “N” 

Содержание SF6 AREACHECK P2

Страница 1: ...Unrivalled Gas Detection ionscience com Register your instrument online to receive your extended Warranty SF6 P2 Instrument User Manual V1 2...

Страница 2: ...y Thank you for purchasing your Ion Science instrument The standard warranty of your instrument can be extended to up to two years To receive your extended warranty you must register your instrument o...

Страница 3: ...inie 73 23 EWG erg nzt durch 93 68 EWG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte...

Страница 4: ...rtup 7 Measuring interval 7 Indicator lights 7 Lamp test and manual reset 8 Configuration via the PC interface 9 Connecting to the host computer 9 User parameters 9 Changing the sensor unit 10 O2 sens...

Страница 5: ...6 ionscience com Installation Location SF6 AREACHECK P2 is designed for indoor wall mounting Since SF6 tends to sink towards the ground in undisturbed air we suggest to place it approx 50 60 cm above...

Страница 6: ...plies in case of power failure and whenever the instrument is taken into setup mode via the host computer interface via the RS232 connector The leftmost 6 position clamp block is provided for network...

Страница 7: ...SF6 P2 Instrument User Manual V1 2 Unrivalled Gas Detection Page 6 of 16 ionscience com Network diagram...

Страница 8: ...tomatically return to full sensitivity Since a reset is also performed when the instrument is taken into setup mode via the host computer interface the above also applies in this case After Power up o...

Страница 9: ...s reflect the last sampling period s result in terms of alarm levels as designated On instruments equipped with the optional oxygen sensor these lights have the following additional functions The two...

Страница 10: ...ard configuration also known as 9600 8 N 1 Further select none for the handshake protocol and TTY as emulation if available The option append CR for received lines should be checked for a more readabl...

Страница 11: ...ensor Fit the replacement SmartSensor Power up the P2 insert fuse observing the hints below NOTE To precisely determine SF6 concentration SF6 AREACHECK P2 requires a reference value representing clean...

Страница 12: ...easonable safety margin sensor life is fixed at one year no matter if shelfed or used this is due to the nature of galvanic cells in which the chemical reaction goes on at all times as long as oxygen...

Страница 13: ...500 ppm O2 1 Mains power 100 240VAC 50 60Hz Power consumption 18 VA Operating temperature range 5 C 45 C Storage temperature range 20 C 60 C Operating humidity range 10 90 non condensing Max load rel...

Страница 14: ...that your Extended Warranty Period has been activated and processed Full details along with a copy of our Warranty Statement can be found by visiting www ionscience com Service Ion Science is pleased...

Страница 15: ...w ism d de Email info ism d de Italian Office ION Science Italia Via Emilia 51 c 40011 Anzola Emilia Bologna ITALY Tel 39 051 0561850 Fax 39 051 0561851 Web www ionscience com it Email info ionscience...

Страница 16: ...V1 1 Layout and font update 28 01 13 SF6 P2 V1 1R Logo only 01 08 2017 SF6 P2 V1 2 Design update 05 05 2020 Disclaimer Information in this manual document is subject to change without notice and does...

Отзывы: