Tiger
LT
PC software
The Snapshots Screen
A snapshot records the settings and calibration data on your TIGER
LT
at any particular point in time. The
‘Snapshots’ screen displays a list of all those stored on your PC.
Click ‘Delete’ to remove a selected snapshot.
To restore the stored settings to your TIGER
LT
, first ensure that your instrument is running correctly and
connected to your PC as detailed previously. Ensure that your instrument is in normal Survey mode, is not in
an alarm condition and that no data logging or Health and Safety readings are being calculated. Click
‘Restore’ against the relevant snapshot
Be aware that this process will replace all setup and calibration files
On the restore screen click ’Yes’. When complete, click ‘Close’ and restart your TIGER
LT
. Your TIGER
LT
will
now be restored to the settings and calibration data stored at the time of that snapshot.
The snapshot menu also allows review of stored data when an instrument is not attached.
Expand the menu until the latest or relevant snapshot is displayed. Double-click on the snapshot and all the
data stored in that snapshot can be accessed.