ion science GasCheck G 1 Скачать руководство пользователя страница 19

                                                                                                    

      GasCheck G Instrument User Manual V2.4

  

 

 

Unrivalled Gas Detect

ion.

   

                 

Page 19 of 26

 

                                           

ionscience.com

 

 

Detecting leaks 

 

 

 

 

 

 

 

 

 

 

 

 

Over view 

Hold the GasCheck G in one hand and draw the 
instrument probe along the area to be checked at 
approximately 25mm per second. When a leak is 
detected retrace the route of the probe at a slower 
rate until the leak is located. Once located hold the 
instrument over the leak until the measurement 
stabilises. The reading can be logged by pressing the 
ENTER key. 
Some readings may be larger than the instrument can 
detect and the measurement units will be replaced by 
'99999' on the display, in which case use the UP or 
DOWN keys to adjust the instruments sensitivity.  
When detecting very small leaks or when detecting 
gasses less sensitive, a rate of 10mm per second may 
be required. 
 
 
 
The outer grey probe cover can be removed to allow better access to restricted areas, if the outer probe is removed 
the following points should be noted:- 
 

 

Avoid bending the inner probe as this will affect the GasCheck G's accuracy. 

 

Avoid placing the probe in liquid or dirt as the probe can become blocked. 

 

Care should also be taken to ensure the brass sensor housing component remains at a constant temperature. 
Avoid touching the brass sensor block with fingers. 

 

GasCheck G can detect changes in Humidity and Carbon dioxide, therefore avoid breathing on the probe. 

 

The thin white tube that covers the inner probe should not be removed, this tube ensures a 1mm gap is 
maintained between the probe and the surface being tested. 

 
 

 

 

Содержание GasCheck G 1

Страница 1: ...Register your instrument online to receive your extended Warranty GasCheck G Instrument User Manual V2 4 Unrivalled Gas Detection ionscience com...

Страница 2: ...urchasing your Ion Science instrument The standard warranty of your instrument can be extended to up to two years To receive your extended warranty you must register your instrument online within one...

Страница 3: ...c Compatibility EMC Standards BS EN 61010 1 2010 Safety requirements for electrical equipment for measurement Control and laboratory use Part 1 General requirements BS EN ISO 9001 2015 Quality managem...

Страница 4: ...Battery status 12 Backlight 12 Gas measurement 12 Sound indication 12 Bar graph 12 Peak hold 12 Functions 13 Contrast 13 Backlight 13 Sound 13 Peak hold 13 Calibration 13 Battery selection 14 View Dat...

Страница 5: ...ay not necessarily be self evident and consequently equipment must be regularly inspected and maintained Ion Science recommends that personnel responsible for equipment use institute a regime of regul...

Страница 6: ...used include Quality assurance Testing seal integrity after product manufacture Laboratory applications Detection of leaks from mass spectrometers and chromatograph equipment Industrial Leaks from ga...

Страница 7: ...y case against the list below before use Should the instrument or any accessory appear damaged or missing then contact the instrument supplier for advice before use Item Qty Description Check 1 1 GasC...

Страница 8: ...to air will affect the rate at which heat transmits from the bead to the block these rates of change are measured and displayed as leak rates The diagram below shows the basic functionality of the Gas...

Страница 9: ...itting new ones unclip the battery holder from the instrument by prising apart the PP3 type clip shown in the illustration Once the battery holder is disconnected from the instrument the batteries can...

Страница 10: ...displays it s running screen and is ready for use ENTER This key is used to select functions and accept settings after a parameter has been adjusted ZERO Pressing the Zero key Zeros out background rea...

Страница 11: ...rious settings should be set and adjusted to suit the application Pressing the Esc key while viewing this Running screen will access the instruments Main Menu Repeated pressing of the Esc returns back...

Страница 12: ...sted Backlight The backlight symbol shows the status of the backlight even in bright daylight Gas measurement The GasCheck G2 and G3 measure the leak rate of gas leakage in which ever units are select...

Страница 13: ...symbol on the main running screen shows the status When the backlight is ON it significantly reduces battery life Backlight OFF 40hrs of use Backlight ON 20hrs of use Sound 1 2 3 The GasCheck G has a...

Страница 14: ...sc key to exit to the Main menu Units 3 GasCheck G instruments have the capability of displaying detected readings in a variety of units ppm cc sec mg m3 and g yr Use the Up and Down keys to select th...

Страница 15: ...F6 2 2 E 5 cc sec 0 000022 cc sec Carbon dioxide CO2 4 0 E 5 cc sec 0 000040 cc sec Methane CH4 2 9 E 5 cc sec 0 000029 cc sec Argon Ar 3 5 E 5 cc sec 0 000035 cc sec Oxygen O2 2 9 E 4 cc sec 0 000290...

Страница 16: ...ts Gas table Upgrade G1 X X X X G2 X X X X X X G3 X X X X X X X X X To upgrade a GasCheck G simply purchase an upgrade code from Ion Science Ltd enter the 12 digit number in the upgrade screen using t...

Страница 17: ...nd When a leak is detected the bar graph will start to fill and the frequency of the audio output will increase this will decrease as the probe moves away from the leak Return the probe to the suspect...

Страница 18: ...be so the spanner fits over the brass nut unscrew the probe counter clockwise direction 4 Withdraw the spanner and probe assembly To refit the short or the long probe follow the steps above but in rev...

Страница 19: ...struments sensitivity When detecting very small leaks or when detecting gasses less sensitive a rate of 10mm per second may be required The outer grey probe cover can be removed to allow better access...

Страница 20: ...G automatically selects the type of Calibration media based on which units the instrument already has selected If the GasCheck G is used to detect leaks measured using ppm or mg m3 units then the inst...

Страница 21: ...ess the ENTER key the instrument will Zero this will take a few seconds but once complete a Ready message will appear 4 Insert the instruments probe into the CalChecks outlet and press the ENTER key T...

Страница 22: ...5 x 2 0 Weight 0 45 kg 1 0 lb Ingress ratings IP20 Length of long probe 300 mm 12 Detection type Thermal conductivity Detection time to T90 1 sec Short probe 9 sec long probe Clear down time 1 sec Sho...

Страница 23: ...k G Instrument User Manual V2 4 Unrivalled Gas Detection Page 23 of 26 ionscience com Accessory Item Number CalCheck A 21500 specifying gas leak rate and pressure 0 0005 cc sec Helium Calibration kit...

Страница 24: ...month of purchase Terms and Conditions apply You will then receive a confirmation email that your Extended Warranty Period has been activated and processed Full details along with a copy of our Warran...

Страница 25: ...Email info ism d de Web ism d de French Office Ion Science France 41 Rue des Etoiles 83240 Cavalaire sur Mer FRANCE Tel 33 613 505 535 Web www ionscience com fr Email info ionscience fr China Office I...

Страница 26: ...N A GasCheck G V2 0 Pages 2 4 updated to correct Quality Management System Page 4 Responsibility of use updated and Legal Notice added Page 19 Contact details moved here P 03 03 11 V1 31 N A GasCheck...

Отзывы: