Workflow for overhead examination
6.1.1
Connect the mount adaptor to the baseplate
Align the two hubs on the mount adaptor to the holes in the
base plate and push up so that the hubs are now in the two
groves allowing lateral motion and rotation.
6.1.2
Positioning the base plate
The dStream interface, (d) is on the tabletop and plugged into the
FlexConnect socket.
Place one short mattress (SM), one long mattress (LM) and the T-
shape (TM) mattress on the tabletop.
If the patient is tall, also place the head-/leg support on the
tabletop to support the patient's feet.
Note
Verify that the rail is in the table groove and the base is placed securely.
6.1.3
Attach the coil to the base plate
Attach the coil to the mount adaptor, with the cable pointing
to the head end.