Instrument Error Codes
Introduction
This section describes the error codes displayed by the iPrep
™
Instrument when
it encounters a problem.
1.
Make a note of the error code including the line number and contact
Technical Support (page 45). Note that the error codes listed in the table
below cannot be fixed by the user. A service engineer will be send to
resolve the error.
2.
After recording the error code, press
ESC
to return to the Main screen.
3.
Press
1
to select the Man screen and return the tips to the original position
by pressing
2
as described on page 33. Remove the card and turn
OFF
the
instrument.
If the protocol is interrupted by an error, the protocol cannot be resumed from
the interrupted point. You need to move the axis to the original position, solve
the error before starting the protocol again from the beginning.
Code Problem
Code Problem
10
Failed return to origins, protocol can’t run
143
Does not match delimited character
11
Limit error, protocol can not run
144
Unexpected number of character strings
behind
12
Failed to return to Z Axis, protocol in run
150
Invalid parentheses position
13
Failed to return to P axis, protocol in run
151
Invalid big parentheses position
14
Failed to return to M axis, protocol in run
160
Does not match variable/conversion table
15
Failed to return to Y axis, protocol in run
161
Does not match with system variable with ( )
16
Z axis limit error, protocol in run
170
Assigned variable is already registered in VT
17
/
171
Invalid variable name to be registered
18 /
180
Non
numeric
data
19
Y axis end limit, protocol in run
181
Set Value of system variable is out of range
20
Z axis time out, protocol in run
190
Not variable and constant
21
P axis time out, protocol in run
200
Invalid " " position
22
M axis time out, protocol in run
210
Label overlapped
23
Y axis time out, protocol in run
211
Invalid label name to be registered
24
Open door in motion
220
VP table Number has not yet been assigned
25
Abnormal input from bottom sensor in
motion
221
Overlapped VP table data
26
Failed to initialize heating block
232
Read data smaller than the first in VP table
27
Failed to initialize motion control board
233
Too much data in VP table
110
System error; (Assigned greater than 10).
250
Argument is not correct
111
Script buffer overflowed
251
Number of argument is too much
Continued on next page