Using the iPrep
™
Purification Instrument,
Continued
Manual Screen
To manually move the axis or return the tip to the origin after a manual stop:
1.
Press
1
to go to the
Manual
screen from the Main screen.
2.
From the Manual screen, you can select
1
to move all axes to the original
positions,
2
to return the Tips to Tip Holders and move all axes to original
positions, or
3
to clean.
3.
Depending on when the protocol was stopped or cancelled,
•
Press
2
to return the Tips to Tip Holders, if tips need to be returned to the
holders. The instrument also moves all axes to the original position.
•
Press
1
to go to the
ORG
display to move all axes to the origin, if the tips
do not need to be returned to the holders.
Once in the
ORG
screen, you can individually move each axis to the origin
by pressing
1
,
2
,
3
,
4
, respectively, or press
0
to return all axes to the origin.
4.
The
Clean
option allows you to clean the piercing unit which may get dirty
over time. To clean the piercing unit, press
3
from the Manual screen.
The
Manual
option displayed on the Clean screen allows you to clean the
piercing unit manually using an alcohol wipe. By pressing the
Manual
option,
the piercing unit is lowered so you can access the unit for cleaning.
5.
Press
ESC
to return to Main menu. You are now ready to a run a new
protocol.