InMode Without Vasculaze Operator Manual
Section 12: System Maintenance
79 of 96
2
.
Hold the Handpiece and the Draining Tool over a sink and start pumping air
through the Handpiece until no more water drains out.
3
.
Disconnect the Handpiece from the Water Draining Tool.
Damage to the system or Handpieces by frozen water will not be covered
under System's warranty.
Only Invasix draining mechanical hand pumps (Invasix P/N AT604837x)
could be used to drain Diolaze, DiolazeXL and Lumecca Handpieces. To
order the Water Draining Tool, contact our Service department.
11.2
Replacing the Deionizer Cartridge
For safety, and proper System performance, replace the cooling water with fresh
distilled water only.
To remove the Deionizer Cartridge:
1.
Drain the water from the System following “Draining Water” procedure.
2.
Turn the System Off.
3.
Push the Slide Latch of the Bottle Door down, and remove the Door.
4.
Holding the Bottle cover with one hand, loosen the Bottle by turning it
clockwise.
5.
Disconnect the Deionizer Cartridge from the Water Hoses by pressing the
Release Button on the upper part of the Deionizer (Figure
To install a new Deionizer Cartridge:
1
.
Connect the Water Hose into the Deionizer Filter until is securely locked.
2
.
Remove the Water Bottle from the compartment, pour out the distilled water
from the Bottle according to “Draining Water” procedure and wash it with new
fresh water.
3
.
Fill 1 liter of distilled water into the Bottle and return it to its compartment.
Perform “Filling Water” procedure as described above.