Setup
4 Setup
4.1 Setup/Delivery Inspection
Setup/delivery inspection should be performed by
provider at time of delivery/set up.
Initial adjustments should be made to suit your
personal body structure needs and preference.
Thereafter weekly, monthly and periodic inspections
should be performed by user/attendant between
the six month service inspections. Refer to
7
Maintenance, page 68
.
q
Check all parts for shipping damage. In case of damage,
DO NOT use.
q
Ensure wheelchair rolls straight (no excessive drag or
pull to one side).
q
Ensure clothing guards are secure.
q
Ensure arms are secure but easy to release and
adjustment levers engage properly.
q
Ensure adjustable height arms operate and lock securely.
q
Ensure axle nut or bolt and wheel mounting nuts are
secure on drive wheels.
q
Ensure caster/anti-tipper wheels are free of debris,
and all mounting hardware is secure and not
damaged/missing.
q
Check that cables are routed and secured properly
to ensure that cables do NOT become entangled and
damaged during normal operation of seating system.
q
Ensure proper operation of powered functions
(Example: drive, seating and legrests).
4.2 Adjusting Ultra Low Maxx Systems—General
Information
CAUTION!
Damage to mobility device and accident hazard
It is possible that collisions can occur between
mobility device components due to various
combinations of adjustment options and their
individual settings
– The mobility device is equipped with an
individual, multiply adjustable seating system
including adjustable legrests, armrests, a
headrest or other options. These adjustment
options are described in the following chapters.
They are used to adapt the seat to the physical
requirements and the condition of the user.
When adapting the seating system and the seat
functions to the user, ensure that no mobility
device components collide.
60101832-A
31
Содержание Ultra Low Maxx
Страница 83: ...Notes ...