Invacare® TDX® SP2 Series
Invacare strongly recommends battery installation
and replacement be performed by a qualified
technician.
Read the installation instructions noted on the
battery and in the manual.
Ensure there is no battery acid in the bottom or
around the battery box or on the sides of the
battery(ies). Neutralize battery acid with baking soda
if found.
Clean battery tray and batteries prior to installation.
Use battery lifting strap when available.
Keep batteries in an upright position and avoid
tipping.
Use only deep cycle gel cell battery.
Use box wrench when tightening clamps. Do Not
wiggle the battery terminals/posts when tightening.
7.2.12 Removing the Batteries from the Wheelchair
CAUTION!
Risk of Damage
Cleaning or maintenance may cause damage to
carpeting or flooring.
– Place the wheelchair in a well ventilated
area where cleaning or maintenance can be
performed without risk of damage to carpeting
or flooring.
Place two 5-inch blocks under battery box to lift
frame off the ground for ease in performing this
procedure. Make sure the battery box is sitting
flat on the blocks and the wheelchair will not be
wobbly during service procedures. When procedure
is complete remove the blocks from underneath
wheelchair.
22NF Batteries shown
1.
Place the wheelchair in a well ventilated area.
2.
Remove the front riggings. Refer to the Captain’s Seat
User Manual.
3.
Verify the remote On/Off switch is in the Off position
and disconnect by unplugging the cable from the
remote.
4.
Remove the rear shroud and front shroud. Refer to
Removing/Installing the Shrouds, page 70
.