![Invacare Stratos IRC 1720 Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/invacare/stratos-irc-1720/stratos-irc-1720_operators-manual_2082631031.webp)
SECTION 4—CLEANING AND DISINFECTING
Part No 1130202
31
Stratos™Portable
Cleaning the Compressor
CAUTION
Unplug unit before cleaning.
NEVER submerge the nebulizer compressor in water or
the AC/DC adapter or auto adapter, hold under running
water or spray with water or other liquids. This will result
in compressor damage.
If fluid gets into the area around the compressor outlet,
wipe the area off with a dry cloth and let the nebulizer sit
on its front end for five minutes. Then run the compressor
for two minutes with no nebulizer or tubing adapter in
place.
WARNING
DO NOT clean using abrasive cleaners.
1. Make sure the nebulizer is off and is not plugged into any
power source.
2. Clean the casing of the main unit daily by using a soft cloth
moistened with water or a mild detergent.
All manuals and user guides at all-guides.com
all-guides.com