SECTION 2—COMPONENT REPLACEMENT
SOLO
2
™
10
Part No 1164898
SECTION 2—COMPONENT
REPLACEMENT
CAUTION
To
insure
proper
operation
and
avoid
voiding
unit
warranties
only
use
Invacare
supplied
components.
Cleaning the Gross Particle Filter
NOTE:
For
this
procedure,
refer
to
FIGURE 2.1.
NOTE:
The
gross
particle
filter
should
be
cleaned
or
replaced
according
to
the
Transportable
Oxygen
Concentrator
Preventive
Maintenance
Record
on
page 23.
CAUTION
DO
NOT
operate
the
Transportable
Oxygen
Concentrator
without
filters
installed.
DO
NOT
clean
the
filter
with
alcohol
and
alcohol
based
products
(isopropyl
alcohol),
concentrated
chlorine
based
products
(ethylene
chloride),
or
oil
based
products
(Pine
‐
Sol®,
Lestoil®)
or
any
other
harsh
chemical
agents.
Only
use
mild
liquid
dish
detergent
(such
as
Dawn™).
1. Squeeze
thumb
tabs
on
filter
cover
and
remove
from
unit.
2. Lift
out
air
intake
filter.
3. Use
a
vacuum
cleaner
or
wash
filter
with
a
mild
liquid
dish
detergent
(such
as
Dawn
™
)
and
water.
Rinse
thoroughly.
4. Thoroughly
dry
the
filter
and
inspect
for
fraying,
crumbling,
tears
and
holes.
A. Replace
filter
if
damaged.
5. Reinstall
the
filter
cover
by
placing
tabs
in
slots,
push
to
engage
into
place.
NOTE:
Use only Invacare replacement part
number 1156863 (Gross Particle Filter) or
1156861 (Gross Particle Filter and Filter
Cover Kit) on the Transportable Oxygen
Concentrator.
L
FIGURE 2.1
Cleaning the Gross Particle
Filter
Filter
Thumb
Tabs
Filter
Cover
Filter
Cover