SECTION 2—COMPONENT REPLACEMENT

SOLO

2

10

Part No 1164898

SECTION 2—COMPONENT 
REPLACEMENT

CAUTION

To

 

insure

 

proper

 

operation

 

and

 

avoid

 

voiding

 

unit

 

warranties

 

only

 

use

 

Invacare

 

supplied

 

components.

Cleaning the Gross Particle Filter

NOTE:

 

For

 

this

 

procedure,

 

refer

 

to

 

FIGURE 2.1.

NOTE:

 

The

 

gross

 

particle

 

filter

 

should

 

be

 

cleaned

 

or

 

replaced

 

according

 

to

 

the

 

Transportable

 

Oxygen

 

Concentrator

 

Preventive

 

Maintenance

 

Record

 

on

 

page 23.

CAUTION

DO

 

NOT

 

operate

 

the

 

Transportable

 

Oxygen

 

Concentrator

 

without

 

filters

 

installed.

DO

 

NOT

 

clean

 

the

 

filter

 

with

 

alcohol

 

and

 

alcohol

 

based

 

products

 

(isopropyl

 

alcohol),

 

concentrated

 

chlorine

 

based

 

products

 

(ethylene

 

chloride),

 

or

 

oil

 

based

 

products

 

(Pine

Sol®,

 

Lestoil®)

 

or

 

any

 

other

 

harsh

 

chemical

 

agents.

 

Only

 

use

 

mild

 

liquid

 

dish

 

detergent

 

(such

 

as

 

Dawn™).

1. Squeeze

 

thumb

 

tabs

 

on

 

filter

 

cover

 

and

 

remove

 

from

 

unit.

2. Lift

 

out

 

air

 

intake

 

filter.

3. Use

 

a

 

vacuum

 

cleaner

 

or

 

wash

 

filter

 

with

 

a

 

mild

 

liquid

 

dish

 

detergent

 

(such

 

as

 

Dawn

)

 

and

 

water.

 

Rinse

 

thoroughly.

4. Thoroughly

 

dry

 

the

 

filter

 

and

 

inspect

 

for

 

fraying,

 

crumbling,

 

tears

 

and

 

holes.

 

A. Replace

 

filter

 

if

 

damaged.

5. Reinstall

 

the

 

filter

 

cover

 

by

 

placing

 

tabs

 

in

 

slots,

 

push

 

to

 

engage

 

into

 

place.

NOTE:

 

Use only Invacare replacement part 

number 1156863 (Gross Particle Filter) or 
1156861 (Gross Particle Filter and Filter 
Cover Kit) on the Transportable Oxygen 
Concentrator.

L

FIGURE 2.1   

Cleaning the Gross Particle 

Filter

Filter

Thumb 

Tabs

Filter 

Cover

Filter 

Cover

Содержание SOLO2 TPO100B

Страница 1: ...es in this manual MUST be performed by a qualified technician For more information regarding Invacare products parts and services please visit www invacare com SOLO2 Transportable Oxygen Concentrator...

Страница 2: ...BLE TO UNDERSTAND THE WARNINGS CAUTIONS OR INSTRUCTIONS CONTACT A HEALTHCARE PROFESSIONAL DEALER OR TECHNICAL PERSONNEL BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE INJURY OR DAMAGE MAY OCCUR ACC...

Страница 3: ...the Inlet Filter 11 Replacing the Patient Outlet Filter 12 Replacement Parts 12 SECTION 3 CHECKING O2 PURITY 13 SECTION 4 ADMINISTRATIVE SETTINGS SCREEN 14 Accessing Administrative Settings Screen 14...

Страница 4: ...hat normally would not burn are easily ignited and burn with great intensity in oxygen enriched air Failure to observe this warning can result in severe fire property damage and cause physical injury...

Страница 5: ...otected against dripping water in upright position AC and DC power only IPX2 Protected against dripping water in upright and tilted positions battery power only NO AP APG NOT suitable for use in the p...

Страница 6: ...cc delivered per minute CONSERVER BREATH RATE CAPACITY Up to and including 35 BPM without reduction of bolus minute volume CONTINUOUS FLOW SETTINGS 0 5 through 3 0 LPM 0 psi 0 5 LPM increments All se...

Страница 7: ...DC SUPPLY 95 F to 104 35 C to 40 C Unlimited use all settings all modes Unlimited use all settings with pulse mode Limited to 2 0 LPM or less with continuous mode EXTENDED TEMPERATURE RANGE USING BAT...

Страница 8: ...e The Risk Of Burns Electrocution Fire Or Injury To Persons DO NOT come in contact with the concentrator while wet DO NOT place or store product where it can drop into water or other liquid DO NOT rea...

Страница 9: ...omagnetic emissions permitted by the above standards To determine if the emissions from the Transportable Oxygen Concentrator are causing the interference turn the Transportable Concentrator Off If th...

Страница 10: ...lorine based products ethylene chloride or oil based products Pine Sol Lestoil or any other harsh chemical agents Only use mild liquid dish detergent such as Dawn 1 Squeeze thumb tabs on filter cover...

Страница 11: ...ps screw driver remove the retaining screw securing the inlet filter access door to unit DETAIL B 3 Remove inlet filter access door from unit 4 Locate existing inlet filter and remove DETAIL C 5 Insta...

Страница 12: ...filter and turn counter clockwise to loosen DETAIL A 3 Remove patient outlet filter and o ring from unit DETAIL B 4 Install new patient o ring onto unit 5 Install patient outlet filter A Hand tighten...

Страница 13: ...oncentrator Preventive Maintenance Record on page 23 1 Turn the unit On by pressing the On Off button 2 Adjust flow to 3 0 LPM continuous refer to owner s manual allow to run for a minimum of 8 minute...

Страница 14: ...Hours NOTE For this procedure refer to FIGURE 4 1 1 Enter administrative settings screen DETAIL B Refer to Accessing Administrative Settings Screen on page 14 for this procedure 2 Locate the Patient H...

Страница 15: ...ocedure 2 Locate the DC Power Limit setting DETAIL B A Press the Up Increase or Down Decrease button to highlight DC Power Limit setting B Press the Return Highlight button to relocate highlight bar t...

Страница 16: ...to the adjustable setting on right side of screen 3 Press the Up Increase or Down Decrease button to toggle between ON or OFF 4 Save changes by pressing the Return Highlight button to relocate highli...

Страница 17: ...ery bars on battery symbol that correspond to the battery and the charging icon if battery is not at 100 DETAIL B FIGURE 5 1 Functional Test 4 Turn the Transportable Oxygen Concentrator on by pressing...

Страница 18: ...ximately 2 3 seconds Verify unit starts and allow unit to run for 30 seconds 15 Turn unit off to complete test Checking Outlet HEPA Filter NOTE Recommended to be replaced between patients 1 Turn unit...

Страница 19: ...r O2 Sensor Fail Alarm This alarm centers on failure of the O2 sensor This alarm has no corrective action available to either the patient or the provider Accessing Alarm Codes To access the Alarm code...

Страница 20: ...Temp HI LO Check ambient temp and blockage of air inlet outlet 2304 Warning Low O2 Purity Check ambient temp and blockage of air inlet outlet 2560 Alarm Low O2 Purity Check ambient temp and blockage...

Страница 21: ...error 2832 Alarm ric Event queue full 2833 Alarm ric Event queue empty 2834 Alarm ric Time out heap full 2835 Alarm ric Time out not found 2836 Alarm ric Infinite null transitions 2837 Alarm ric Too...

Страница 22: ...river i2c response time out 2861 Alarm Wait for V12 off time out 2862 Alarm Wait for V12 on time out 2863 Alarm Gpio exp driver wait to acquire i2c time out 2864 Alarm Battery controller i2c response...

Страница 23: ...days 4 380 hours are equivalent to usage 24 hours per day 7 days per week for 6 months 26 280 hours are equivalent to usage 24 hours per day 7 days per week for 3 years Record Date Of Service Record...

Страница 24: ...erland Tel 41 22 354 60 10 Fax 41 22 354 60 11 Canada 570 Matheson Blvd E Unit 8 Mississauga Ontario L4Z 4G4 Canada 800 668 5324 All rights reserved Trademarks are identified by and All trademarks are...

Отзывы: