Usage

6.3.3

Flowrate

WARNING!
Risk of Injury

To avoid injury from excess oxygen or a deficit of
oxygen:

DO NOT change the L/min setting on the flowmeter

unless a change has been prescribed by your physician
or therapist.

DO NOT set the flow greater than 10 L/min.

1.

Turn the flowrate knob

A

to the setting prescribed by your

physician or therapist.

To properly read the flowmeter

B

, locate the

prescribed flowrate line on the flowmeter. Next, turn
the flow knob until the ball

C

rises to the line. Now,

center the ball on the L/min line prescribed.

2.

If the flowrate on the flowmeter ever falls below 1 L/min for
more than about one minute, the Potential Obstruction Alert
will be triggered. This is a rapid beeping of the audible alarm.
Check your tubing or accessories for blocked or kinked tubing
or a defective humidifier bottle. After rated flow is restored to 1
L/min, the Potential Obstruction Alert will turn off.

A Potential Obstruction Alert indicates a condition that
may be associated with a partial or complete obstruction
of oxygen output.

The use of some accessories and the HomeFill compressor
will deactivate the Potential Obstruction Alert.

6.3.4

Units without SensO

2

Concentrator may be used during the initial start warm-up
time (approximately 30 min). O

2

purity will build to a

maximum during this period.

Light

Symbol

Explanation

A

GREEN

light

Normal Operation

B

RED

light

Continuous Audible Alarm

Sieve-GARD™Compressor

Total unit shutdown. Immediately
switch to backup oxygen supply
and call supplier.

C

Call Supplier

1118353-M-00

21

Содержание PLATINUM 10L

Страница 1: ...Platinum 10L Oxygen Concentrator IRC10LX en HomeFill System Compatible User Manual This manual MUST be given to the user of the product BEFORE using this product read this manual and save for future r...

Страница 2: ...on from Invacare Trademarks are identified by and All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted Making Life s Experiences Possible is a tra...

Страница 3: ...2 Select a Location 17 6 3 Setting Up the Concentrator 18 6 3 1 Setting Up the Humidifier 18 6 3 2 Turning the Concentrator On 20 6 3 3 Flowrate 21 6 3 4 Units without SensO2 21 7 Maintenance 22 7 1 S...

Страница 4: ...avoided may result in property damage or minor injury or both Gives useful tips recommendations and information for efficient trouble free use Refer to the following table for the meaning of symbols m...

Страница 5: ...ualified technician 1 2 Intended Use The intended function and use of the Invacare Platinum Oxygen Concentrator is to provide supplemental oxygen to patients with respiratory disorders by separating n...

Страница 6: ...For Use The intended function and use of the Invacare Platinum Oxygen Concentrator is to provide supplemental oxygen to patients with respiratory disorders by separating nitrogen from room air by way...

Страница 7: ...Safety 2 1 Label Locations A Serial Number Label is located on the resonator intake assembly B C Specification and Double Insulation Label is located on the back of the concentrator at the base 111835...

Страница 8: ...Invacare Platinum 10L Oxygen Concentrator IRC10LX D 8 1118353 M 00...

Страница 9: ...age From Fire Textiles oil or petroleum substances grease greasy substances and other combustibles are easily ignited and burn with great intensity in oxygen enriched air and when in contact with oxyg...

Страница 10: ...ed with this device You should contact your local home care supplier for recommendations on which of these devices will be best for you They should also give you advice on the proper usage maintenance...

Страница 11: ...Make sure concentrator is off when not in use WARNING Risk of Injury or Damage Invacare oxygen concentrators are specifically designed to minimize routine preventive maintenance To prevent injury or...

Страница 12: ...a reasonable protection against electromagnetic interference in a typical medical installation Other devices may experience interference from even the low levels of electromagnetic emissions permitted...

Страница 13: ...compressor IOH200 The HomeFill outlet fitting I is to be used only for filling oxygen cylinders with the HomeFill home oxygen compressor The outlet fitting does not affect concentrator performance Re...

Страница 14: ...Invacare Platinum 10L Oxygen Concentrator 3 2 Pneumatic Diagram 14 1118353 M 00...

Страница 15: ...nula mask tubing humidifier etc used to deliver oxygen to the patient need to include a means to reduce the propagation of fire in the accessories for the safety of the patient and others If a commerc...

Страница 16: ...he carton Unless the oxygen concentrator is to be used IMMEDIATELY leave concentrator in its packaging for storage until use is required 5 2 Inspection 1 Inspect examine exterior of the concentrator f...

Страница 17: ...imal performance Locate and position the concentrator in a well ventilated space so that the air intake and the air exhausts are not obstructed NEVER block the air openings of the concentrator or plac...

Страница 18: ...t if input and output connections are reversed When using tubing connections longer than 7 ft 2 1 m in length position the humidifier as close to the patient as possible to allow for maximum humidific...

Страница 19: ...ure prior to use CAUTION Risk of Damage Confirm the humidifier cap is not cross threaded on the humidifier bottle 3 Insert a flathead screwdriver in the plate groove C on the top edge of the filter ac...

Страница 20: ...s longer than 7 ft 2 1 m in length position the humidifier as close to the patient as possible to allow for maximum humidification output 7 Attach the patient s nasal cannula supply tube to the humidi...

Страница 21: ...be triggered This is a rapid beeping of the audible alarm Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle After rated flow is restored to 1 L min the Pot...

Страница 22: ...et DANGER Risk of Injury or Damage Liquid will damage the internal components of the concentrator To avoid damage or injury from electrical shock Turn Off the concentrator and unplug the power cord be...

Страница 23: ...tor without the filter installed or with a dirty filter There are two cabinet filters located on each side of the cabinet 1 Remove the filter A and clean as needed Environmental conditions that may re...

Страница 24: ...r Only qualified personnel should perform cleaning and disinfection of the oxygen concentrator and accessories between patients Follow these instructions to eliminate possible pathogen exchange betwee...

Страница 25: ...EN PATIENTS UNITS WITHOUT SensO2 Every 6 months of continuous use Equivalent to 4 380 hours Check Oxygen Concentration green indicator light Clean Replace Cabinet Filter s Refer to Cleaning the Cabine...

Страница 26: ...individual product marking for disposal 8 2 Wear and Tear Invacare reserves the right to ask for any item back that has an alleged defect in workmanship See Warranty that shipped with the product for...

Страница 27: ...nit overheating due to blocked air intake 2 Insufficient power at outlet 3 Internal repairs required 1 Do one or both of the following a Remove and clean cabinet filters b Move oxygen concentrator at...

Страница 28: ...evels 87 to 95 6 at 2 L to 10 L min Concentration levels achieved after initial warm up period approximately 30 minutes Maximum Outlet Pressure 9 psi 0 5 psi 62 0 kPa 3 45 kPa Flow Range 2 L to 10 L m...

Страница 29: ...0 lbs 2 lbs 27 2 kg 1 kg Operating Ambient Temperature and Humidity 50 F to 95 F 10 C to 35 C at up to 60 relative humidity Cabinet Impact Resistant flame retardant plastic cabinet that conforms to UL...

Страница 30: ...t 15 m of Crush Proof Tubing DO NOT pinch Time of Operation Up to 24 hours per day Recommended Storage and Shipping Temperature 20 F to 150 F 29 C to 65 C at up to 95 relative humidity Measurement unc...

Страница 31: ...ormal wear and tear or shipping charges Invacare Corporation s sole obligation and your exclusive remedy under this warranty shall be limited to such repair or replacement Routine maintenance items su...

Страница 32: ...IVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN INVACARE SHALL NOT BE...

Страница 33: ...Notes...

Страница 34: ...Notes...

Страница 35: ...Notes...

Страница 36: ...Tel 440 329 6000 Tel 800 832 4707 Technical Services Tel 440 329 6593 Tel 800 832 4707 www invacare com Manufacturer Invacare Corporation 2101 E Lake Mary Blvd Sanford FL 32773 Tel 407 321 5630 111835...

Отзывы: