![Invacare Perfecto2 IRC5PAW Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/invacare/perfecto2-irc5paw/perfecto2-irc5paw_operators-manual_2082543028.webp)
SECTION 8—TROUBLESHOOTING GUIDE
Perfecto
2
™Series
28
Part No 1163145
YELLOW or RED
light Illuminated
Alarm:
Continuous
On RED light only
* Only applies to
IRC5PO2AW &
IRC5PO2VAW
models
1. Low oxygen purity.*
1. Clean or Replace filters.
2. Kinked or blocked
tubing, cannula or
humidifier.*
2. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back On.
3. Flowmeter set at 0.5
L/min *.
3. Check flowmeter is set to
1.0 L/min or more. Refer to
TYPICAL PRODUCT
PARAMETERS on page 13.
4. Unit overheating due
to blocked air intake.
4a. Remove and clean cabinet
filters.
4b. Move concentrator at
least three inches from walls,
draperies, and furniture.
5. Insufficient power at
outlet.
5. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
6. Internal repairs
required.
6. Call Service Provider.
GREEN light with
YELLOW Light
Flashing
1. Internal repairs
required.
1. Call Service Provider.
Alarm:
Rapid
Beep....Beep...
Beep...Beep
LOW FLOW ALARM
1a. Kinked or blocked
tubing, cannula or
humidifier.
2. Flowmeter set at 0.5
L/min.
1a. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back ON.
2. Flowrates less than 1 L/min
are not recommend.
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
Содержание Perfecto2 IRC5PAW
Страница 26: ...SECTION 7 MAINTENANCE Perfecto2 Series 26 Part No 1163145 FIGURE 7 2 Preventive Maintenance Record...
Страница 32: ...Perfecto2 Series 32 Part No 1163145 NOTES...
Страница 59: ...SECTION 7 ENTRETIEN Part No 1163145 59 Perfecto2 S rie FIGURE 7 2 Rapport de maintenance pr ventive...
Страница 92: ...SEZIONE 7 MANUTENZIONE Serie Perfecto2 92 Part No 1163145 FIGURA 7 2 Verbale di manutenzione preventiva...
Страница 126: ...ABSCHNITT 7 WARTUNG Perfecto2 Series 126 Part No 1163145 ABBILDUNG 7 2 Aufzeichnungen ber vorbeugene Wartung...
Страница 161: ...CZ 7 KONSERWACJA Part No 1163145 161 Serie Perfecto2 FIGURE 7 2 REJESTR KONSERWACJI PREWENCYJNEJ...
Страница 193: ...ST 7 DR BA Part No 1163145 193 Perfecto2 Series OBR ZEK 7 2 Z znam preventivn dr by...