Invacare MicroAir MA90Z Скачать руководство пользователя страница 23

4 OPERATION

Part No 1148139

23

Invacare

®

 MA90Z Series Mattresses

Automatic Wireless Fowler 

Manual Operation of the Fowler

1.

Press the Select button (

) to move the pointer on the LCD to Function.

2.

Use the Firm and Soft buttons until Fowler displays.

About Automatic Wireless Mode

When the bed articulates to 45°, the transmitter in the head of the mattress signals the control unit to add 25% more air into the mattress to prevent 

the patient from bottoming out.

Programming the Automatic Wireless Fowler Transmitter

1.

Ensure the transmitter is inside the mattress pocket located on the right side of the mattress at the head end.

2.

Ensure the mattress is in a flat position.

3.

Ensure the control unit is plugged in, and the LCD displays STAND BY.

4.

Perform one of the following:

MA90Z Power Unit - Perform the following steps:

i.

Press and hold the Turn Angle and Max Inflate buttons.

ii.

Raise the bed to 45°.

When this mode is activated, the control unit increases the pressure in the mattress to prevent the patient from bottoming 
out.

The Fowler LED is lit and L displays in the LCD Display.

The control unit will beep and will return to Standby

.

Содержание MicroAir MA90Z

Страница 1: ...42 Lateral Rotation with Alternating Pressure and On Demand Low Air Loss Bariatric 42 MA90ZB48 Lateral Rotation with Alternating Pressure and On Demand Low Air Loss Bariatric 48 Invacare MA95Z Series...

Страница 2: ...rights reserved Republication duplication or modification in whole or in part is prohibited without prior written permission from Invacare Trademarks are identified by and All trademarks are owned by...

Страница 3: ...lacement System 14 Installing the Side Rails 16 Installing the Power Unit 16 Connecting the Hose 16 Connecting the Power Cord 18 Using the Front Panel 19 Powering Up the System 26 Placing the Patient...

Страница 4: ...hazardous situation which if not avoided will result in death or serious injury WARNING Warning indicates a potentially hazardous situation which if not avoided could result in death or serious injury...

Страница 5: ...ectly to Invacare at the address on the back cover Provide dealer s name address model number and the date of purchase indicate nature of the defect and if the product is serialized indicate the seria...

Страница 6: ...cause injury or death Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment Follow the manufacturer s instructions Monitor patient frequently Read and understand the Owner s...

Страница 7: ...5Z power unit is also used on MA95ZB42 and MA95ZB48 systems MA90Z MA95Z INPUT VOLTAGE AC 110 V INPUT FREQUENCY 60 Hz CURRENT 1 A 2 A MAXIMUM POWER CONSUMPTION 30 10 W 150 10 W CIRCUIT PROTECTION Dual...

Страница 8: ...MAXIMUM FLOW TIMER 30 minutes 15 minutes SUPPORT SURFACE INFLATION TIME 5 10 minutes 25 60 seconds PATIENT COMFORT CONTROL PRESSURES SOFT PRESSURE FIRM PRESSURE 8 4 mmHg 32 4 mmHg ROTATION TIME 10 20...

Страница 9: ...ROL UNIT CONNECTION Quad Flow Four Single Quick Coupling Triple Flow Two and One Single Magnetic Quick Coupling CONTROL UNIT PACKAGING 1 piece box AIR FILTER None Charcoal Air Filter with Fire Retarda...

Страница 10: ...professional dealer or technical personnel before attempting to use this equipment otherwise injury or damage may occur ACCESSORIES WARNINGS Invacare products are specifically designed and manufacture...

Страница 11: ...islation Controls on the footboard may be obstructed by the power unit on a few bed frames It may be necessary to relocate the power unit Refer to Installing the Power Unit on page 16 Check that air h...

Страница 12: ...apment Variations in bed rail dimensions and mattress thickness size or density could increase the risk of entrapment Visit the FDA website at http www fda gov to learn about the risks of entrapment R...

Страница 13: ...mable anesthetics Oxygen There is a possible fire hazard when used with oxygen administering equipment other than nasal mask or half bed tent type The oxygen tent should NOT extend below mattress supp...

Страница 14: ...other brands Otherwise damage to the system may occur Contact your supplier to get the correct replacement if needed CAUTION DO NOT strap the mattress to the bed frame at the head and foot ends Secur...

Страница 15: ...frame 4 Use the buckle straps to secure the powered mattress to the bed deck in the following locations Head End Head End Bed Deck Foot End Foot End Bed Deck Center Center of the Bed Frame FIGURE 1 I...

Страница 16: ...vent patient entrapment Follow the manufacturer s instructions Monitor patient frequently Read and understand the Owner s Operator s Manual prior to using this equipment Invacare product manuals are a...

Страница 17: ...ectors onto the control connector until an audible click is heard MA95Z Power Unit Perform the following steps Detail B i Lift and hold the connector cover up ii Push the hose connector onto the contr...

Страница 18: ...f the control unit 3 Plug the other end of the plug into a properly grounded outlet on the wall WARNING DO NOT alter plug to fit a non conforming outlet Instead have an electrician install a properly...

Страница 19: ...NEL Power Button Turn Button Turn Time Button Firm Button Soft Button Turn Angle Button Mode Max Inflate Low Air Loss Button DETAIL A MA90Z POWER UNIT FRONT PANEL Lock Alarm Button Low Pressure LED Po...

Страница 20: ...es Once the unit is plugged in an AMBER LED on the control unit is lit indicating that the system is in STAND BY mode Once the Power button is pressed and released a GREEN LED illuminates indicating t...

Страница 21: ...PULSE Set 10 minute pulse time cycle All of the air cushions in the mattress will remain static at the current pressure settings for 10 minutes then the pressure in the mattress will decrease by 20 f...

Страница 22: ...Soft Button Pressing this button reduces the pressure in the mattress Firm Button Pressing this button increases the pressure in the mattress FOWLER UPRIGHT Fowler Mode The pressure will increase to p...

Страница 23: ...ng the Automatic Wireless Fowler Transmitter 1 Ensure the transmitter is inside the mattress pocket located on the right side of the mattress at the head end 2 Ensure the mattress is in a flat positio...

Страница 24: ...Power Unit 1 Press the Mode button to select the Max Inflate Mode or the Low Air Loss mode Max Inflate Mode In this mode the mattress inflates rapidly to maximum firmness pressurized to 35 6 mmHg A se...

Страница 25: ...e button again Lock Alarm Silence Button 1 Press the Lock Alarm Silence button to select one of the following modes Lock Mode Pressing this button locks out all control unit functions including the On...

Страница 26: ...unit is plugged in the LCD screen displays the Run Time in hours and minutes For example 2 584 hours and 25 minutes displays as 002584 Hours 000025 Minutes To display the Run Time during use 1 Ensure...

Страница 27: ...patient has not bottomed out by performing the following steps A Ensure that the patient is lying flat on his her back in the middle of the mattress B Place four fingers between the air cushions dire...

Страница 28: ...elchair 1 Engage the wheel locks of the bed Refer to the owner s manual provided with the bed 2 Engage the wheel locks of the wheelchair if applicable Refer to the owner s manual provided with the whe...

Страница 29: ...e hose connector Detail B MA95Z Power Unit Lift the connector cover Detail A 2 Pull the hose connector Details B and C from the control unit connectors 3 MA90Z Only Disconnect the RED CPR connector lo...

Страница 30: ...E 5 Preparing for CPR Procedure Connector Cover ControlUnit Connectors MA95Z MA95ZB42 and MA95ZB48 Hose Connector DETAIL A Control Unit RED CPR Connector MA90Z MA90ZB42 and MA90ZB48 Only Mattress DETA...

Страница 31: ...disposable paper towels ii Scrub the area with freshly prepared effective phenolic detergent disinfectant solution WARNING Before cleaning or disassembling the MA90 series check the underside of the m...

Страница 32: ...according to the manufacturer s instructions iv Remove excess water v Set the dryer to the lowest setting below 120 F vi Dry the cover until it is completely dry 9 Perform the following steps to clean...

Страница 33: ...to Cleaning the System on page 31 3 Disconnect the air hose connector from the control unit and allow air to vent from the mattress 4 Gently roll up the mattress with minimal handling and agitation 5...

Страница 34: ...5 MAINTENANCE AND TROUBLESHOOTING Invacare MA90Z Series Mattresses 34 Part No 1148139 FIGURE 1 Changing the Filter Filter Cover Screws...

Страница 35: ...nked or split manifold Unkink manifold or replace split manifold Has power and fuse is good control unit does not come on Send unit back for repair Not alternating solenoid or valve mal function Send...

Страница 36: ...care Corporation USA One Invacare Way Elyria Ohio USA 44036 2125 800 333 6900 Canada 570 Matheson Blvd E Unit 8 Mississauga Ontario L4Z 4G4 Canada 800 668 5324 Part No 1148139 Rev D 1 11 www invacare...

Отзывы: