11
FREQUENCY CHANNEL/BATTERY
CHANGING FREQUENCY CHANNEL
(FIGURE 1)
NOTE: This procedure is for deluxe model beds only.
The frequency of the pendant has been factory set for a spe-
cific frequency. The shipping box and the control box under-
neath the bed will specify the frequency the pendant is set.
If using two (2) twin size as a king size bed, both beds can be
set to the same frequency to operate with one (1) pendant.
1. Raise the head end of the bed to expose the control box.
2. Unplug the bed from the wall outlet.
3. Locate two (2) prong channel switch on the control box.
4. Switch channel to one (1) of the settings in FIGURE 1.
5. Remove the battery cover on the back of the pendant.
6. Match the channel setting in the pendant with the control
box setting determined in STEP 4.
7. Reinstall the battery cover on the back of the pendant.
8. Plug the bed into the wall outlet.
FIGURE 1 - CHANGING FREQUENCY CHANNEL
Back of
Pendant
Two (2) Prong
Channel Switch
CHANNEL SWITCH SETTINGS
A
B
C
D
F
R
E
Q
U
E
N
C
Y
C
H
A
N
N
E
L
FREQUENCY CHANNEL/BATTERY
B
A
T
T
E
R
Y
Back of
Pendant
REPLACING BATTERY (FIGURE 2)
NOTE: This procedure is for deluxe model beds only.
NOTE: Deluxe Model Pendants require one (1) 12V bat-
tery.
1. Remove the battery cover on the back of the pendant.
2. Note existing battery orientation for installation of new
battery.
3. Remove the existing battery from the pendant.
4. Install the new battery into the pendant in the orienta-
tion noted in STEP 2.
5. Reinstall the battery cover on the back of the pendant.
FIGURE 2 - REPLACING BATTERY
12V Battery
Orientation