Invacare® HomeFill® Oxygen System

device operation stops, then the compressor is causing
the interference. In such rare cases, interference may be
reduced or corrected by one of the following measures:

Reposition, relocate, or increase the separation between
the devices.

Connect either line powered device to a different
electrical power circuit.

Contact Invacare for any additional EMC information or EMC
tables for the device environment.

12

1100873-M

Содержание HomeFill Oxygen System

Страница 1: ...Invacare HomeFill Oxygen System Model IOH200 en Compressor User Manual This manual MUST be given to the user of the product BEFORE using this product read this manual and save for future reference ...

Страница 2: ...owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted Platinum is a registered trademark in the U S A Europe and Australia Making Life s Experiences Possible is a registered trademark in the U S A 3 in 1 oil is a registered trademark of American Home Products Corporation WD 40 is a registered trademark of the WD 40 Company Updated versions of this manual are avail...

Страница 3: ...16 4 Usage 17 4 1 Compressor Operation Checklist 17 4 2 Cylinder Prefill Inspection 17 4 3 Connecting the Cylinder to the Compressor 19 4 4 Disconnecting the Cylinder from the Compressor 20 4 5 Turning the Compressor On and Off 21 5 Maintenance 23 5 1 Maintenance Info 23 5 2 Wear and Tear 23 5 3 Service Life 23 5 4 Cleaning the Compressor Filter 24 5 5 Cleaning the Cabinet 24 5 6 Replacing the Fus...

Страница 4: ... avoided could result in death or serious injury CAUTION Caution indicates a potentially hazardous situation which if not avoided may result in property damage or minor injury or both IMPORTANT Indicates a hazardous situation that could result in damage to property if it is not avoided Gives useful tips recommendations and information for efficient trouble free use Symbols on the Product The follo...

Страница 5: ... to be distributed to any other individual for any purpose No other use of this device is indicated WARNING Risk of Death Injury or Damage The Invacare HomeFill compressor was designed to be used only with Invacare oxygen concentrators Use with other types of devices may cause injury death or property damage DO NOT use the Invacare HomeFill compressor with anything other than an Invacare oxygen co...

Страница 6: ...t is intended for single patient use in the indoor homecare environment only and is not to be used for bulk cylinder refilling No specific product knowledge or training is required other than what is contained in this manual to operate the product It is not intended for supplying oxygen therapy during refilling This product is intended to be used by or under adult supervision only after reading an...

Страница 7: ...sor these instructions MUST be followed THE INFORMATION IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE DANGER Risk Of Death Injury Or Damage From Fire Textiles oil or petroleum substances grease greasy substances and other combustibles are easily ignited and burn with great intensity in oxygen enriched air and when in contact with oxygen under pressure To avoid fire death injury or damage DO...

Страница 8: ...h as WD 40 or 3 in 1 Oil If connection difficulty is experienced verify that the cylinder fitting is being inserted straight down into the compressor coupler If continued difficulty is experienced contact your supplier for assistance DANGER Risk of Electric Shock This unit contains moving and high pressure components There are no user serviceable parts inside the cover To avoid electric shock inju...

Страница 9: ... result in personal injury or product damage WARNING Risk of injury or damage The HomeFill is equipped with a high pressure relief valve to ensure the user s safety When activated this safety feature will make an extremely loud noise To avoid injury or damage If this noise occurs turn the unit Off and contact your Home Health Care Provider or Service Representative WARNING Risk of injury or damage...

Страница 10: ...Invacare HomeFill Oxygen System 2 2 Label Locations A B C D E 10 1100873 M ...

Страница 11: ...amage To reduce the risk of injury or product damage from interference with wireless equipment Keep the compressor at least 9 8 ft 3 0 m away from wireless communication equipment such as wireless home network devices mobile phones cordless phones and base stations walkie talkies etc Medical Electrical Equipment needs to be installed and used according to the EMC information in this manual This eq...

Страница 12: ... interference may be reduced or corrected by one of the following measures Reposition relocate or increase the separation between the devices Connect either line powered device to a different electrical power circuit Contact Invacare for any additional EMC information or EMC tables for the device environment 12 1100873 M ...

Страница 13: ...ressor is to be used IMMEDIATELY leave the compressor in its packaging for storage until use of the compressor is required 3 2 Compressor Initial Setup Checklist Dealers ONLY Refer to 2 3 Dealer Information page 11 before performing this procedure Users ONLY Contact your Invacare dealer or Invacare for assistance with the setup procedure and to report unexpected operation or events Before operatin...

Страница 14: ...urces of ignition out of the room in which this compressor is located and away from where oxygen is being delivered Keep the oxygen tubing cord and compressor out from under such items as blankets bed coverings chair cushions clothing and away from heated or hot surfaces including space heaters stoves and similar electrical appliances The compressor should be at least three inches away from walls ...

Страница 15: ...des of compressor 3 5 Installing the Power Cord This product may come with a polarized plug one blade is wider than the other As a safety feature this plug will fit in a polarized outlet only one way If the plug does not fit fully in the outlet reverse the plug If it still does not fit contact a qualified electrician DO NOT attempt to defeat this safety feature 1100873 M 15 ...

Страница 16: ...ction device to remove all power to the unit when the line cord is disconnected 3 6 Connecting the Compressor to the Concentrator 1 Insert one end of the interconnect hose C into the outlet fitting B on the back of the concentrator A 2 Insert the opposite end of the interconnect hose into the inlet fitting D on the side of the compressor The metal tabs E on the concentrator outlet fitting and the ...

Страница 17: ...ition q Examine the indicator lights on the control panel q Disconnect and remove the full cylinder q Push the compressor power switch to the Off O position q If filling another cylinder repeat this checklist 4 2 Cylinder Prefill Inspection WARNING ONLY use cylinders that have the Invacare HomeFill connection and HomeFill label All cylinders MUST be inspected before attempting to fill Otherwise in...

Страница 18: ...r if they exist Dents or dings Arc burns Oil or grease Any other signs of damage that might cause a cylinder to be unacceptable or unsafe for use 2 Examine the cylinder for evidence of fire or thermal damage Evidence includes charring or blistering of the paint or other protective coating or heat sensitive indicator If fire or thermal damage is found replace the cylinder 3 Inspect the cylinder reg...

Страница 19: ...erwise injury or damage may occur DO NOT use a liquid leak detector to test for leaks Fillport covers not shown for clarity 1 Examine the cylinder gauge If the cylinder pressure is less than 1500 psig proceed to STEP 2 For location of the cylinder gauge refer to the cylinder s manual If the cylinder pressure is greater than 1500 psig DO NOT attempt to top off this cylinder It may not fill 2 Set th...

Страница 20: ...der in the area behind the cylinder fillport 6 Position the cylinder in the compressor cradle G 7 Align the cylinder fillport with the connector fillport 8 Pull UP on the outer ring sleeve of the connector fillport while pushing DOWN on the cylinder to couple the cylinder fillport into the connector fillport The cylinder is properly connected when an audible click is heard 4 4 Disconnecting the Cy...

Страница 21: ...e replaced after filling and whenever not in use If either of the fillport covers are missing contact your healthcare provider before using 6 Place the fillport covers onto the connector fillport G and cylinder fillport H When using the post valve type cylinder attach an oxygen regulator per manufacturers instructions 7 Examine the cylinder gauge to ensure cylinder is full needle pointing into GRE...

Страница 22: ...r to 4 3 Connecting the Cylinder to the Compressor page 19 4 Push the power switch A on the control panel to the On I position 5 Examine the control panel B The following sequence of events should occur a 0 3 minutes The O2 Below Normal YELLOW light C will be lit while the compressor warms up b After three minutes The FILLING GREEN light D will be on while the cylinder is filling c The FULL GREEN ...

Страница 23: ... See the warranty in this manual for specific warranty information Refer to the procedures in this User Manual for proper preventative maintenance schedule and use of the product This is just a general guideline and does not include items damaged due to abuse and misuse Useful life is dependent upon many factors Normal wear and tear items and components for this product are listed below All types ...

Страница 24: ...ure at least once a week 1 Unplug the compressor A Grasp the plug not the cord 2 Remove the filter B 3 Clean the filter with a vacuum cleaner C or wash in warm soapy water and rinse thoroughly D 4 Dry the filter thoroughly before reinstallation 5 5 Cleaning the Cabinet DANGER Risk of Electric Shock This unit contains moving and high pressure components There are no user serviceable parts inside th...

Страница 25: ...is for an optional spare fuse When replacing the front fuse ensure that there is a spare fuse with the same type and rating in the fuse drawer 5 Install replacement fuse in front fuse D slot Use a 5 mm x 20 mm time lag low breaking capacity 250V 2 Amp IEC 60127 style only Invacare part number 1099820 6 Push fuse drawer back into the power inlet 7 Install power cord into the power inlet on the comp...

Страница 26: ...n symbol is placed on this product to encourage you to recycle wherever possible Follow local governing ordinances and recycling plans regarding disposal of the device or components normally used in operation The device does not generate waste or residue in operation DO NOT dispose of the device in the normal waste stream Any accessories not part of the device MUST be handled in accordance with th...

Страница 27: ... minutes after power is On O2 flowing to user if concentrator is On Cylinder filling stops will restart when O2 90 O2 BELOW NORMAL YELLOW O2 from concentrator is 90 any time after three minute warmup O2 flowing to user if concentrator is On Cylinder is filling FILLING GREEN Compressor is On and cylinder is filling O2 from concentrator is 90 any time after three minute warmup O2 flowing to user if ...

Страница 28: ...nder filling stops Cannot fill another cylinder until compressor is turned Off and On again O2 flowing to user if concentrator is On ATTENTION RED Audible Alarm Compressor is On Cylinder NOT filling Causes Full cylinder attached when unit turned On replace with empty cylinder or Internal failure contact dealer Compressor does not begin to fill or stops filling Unit must be turned Off and On before...

Страница 29: ...our home health care provider or Invacare for service RED Light On 1 Check the cylinder gauge to see if it is full Cylinder full Remove the cylinder and turn the unit Off Perform the 4 1 Cylinder Prefill Inspection on page 20 page 17 with a cylinder that is not full Cylinder not full Proceed to Step 2 2 Check that the cylinder is connected properly 3 Turn the compressor Off 4 Remove the cylinder a...

Страница 30: ...ntact your home health care provider or Invacare O2 below normal YELLOW light stays On 1 Ensure the concentrator has warmed up for at least 20 minutes If it has not been 20 minutes turn the compressor Off until the concentrator warms up If the YELLOW light still remains on proceed to step 2 2 Inspect the connection between the compressor and the concentrator for damage or pinching If damaged repla...

Страница 31: ...low setting refer to the concentrator User manual If the GREEN filling light does not go Off the compressor and or concentrator may need service Contact your home health care provider or Invacare Excessive noise when compressor is On 1 Call home health care provider or Invacare for service Hissing sound from Regulator when flow selector is Off 1 Turn flow selector Off 2 If hissing sounds continue ...

Страница 32: ... interlocking fill fitting inlet fitting and drive motor The compressor operation is controlled by the electronics assembly The motor compressor is allowed to fill a cylinder only when the oxygen concentration of the inlet gas stream is greater than 90 The motor compressor fill operation is completed when the gas in a cylinder reaches the maximum fill pressure of 2000 PSI 13786 kPa The IOH200 oxyg...

Страница 33: ...ep pile carpets and heaters radiators or hot air registers Floor location only Tubing 2 m 7 ft cannula with a maximum 15 m 50 ft of Crush Proof Tubing DO NOT pinch Time of Operation Up to 24 hours per day Recommended Storage and Shipping Temperature 29 C to 65 C 20 F to 150 F at 15 95 relative humidity Environment Smoke and soot free No confined spaces Example No closets Minimum Operating Time 30 ...

Страница 34: ... Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty or 5 000 hours whichever comes first For Model Nos IOH2MD IOH2M6 and IOH2M9 Invacare warrants the HomeFill regulator NOT including the cylinder to be free from defects in materials and workmanship for a period of two years from date of purchase from Invacare or a dealer with a copy of the seller s i...

Страница 35: ...HOUT INVACARE CORPORATION S EXPRESS WRITTEN CONSENT INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS OR TO PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE CORPORATION OR TO PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE CORPORATION S CONTROL THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU ...

Страница 36: ...Elyria Ohio USA 44036 2125 800 333 6900 Technical Services 800 832 4707 www invacare com Manufacturer Invacare Corporation 2101 E Lake Mary Blvd Sanford FL USA 32773 407 321 5630 1100873 M 2018 05 22 1100873M Making Life s Experiences Possible ...

Отзывы: