SECTION 5—INDICATOR LIGHTS
31
89545475293569
INDICATOR LIGHT
COLOUR
COMPRESSOR
STATUS
EFFECT
None
Compressor is Off.
O
2
flowing to user if
concentrator is On.
O
2
Below Normal -
YELLOW
Compressor is On and
warming up (0-3
minutes after power is
On).
Cylinder filling not
started.
O
2
flowing to user if
concentrator is On.
O
2
from concentrator
is <90% (anytime after
three minute warmup).
Cylinder filling stops,
will restart when O
2
>90%.
O
2
flowing to user if
concentrator is On.
Filling - GREEN
Compressor is On and
cylinder is filling. O
2
from concentrator is
>90% (any time after
three minute warmup).
Cylinder is filling
O
2
flowing to user if
concentrator is On.
Full - GREEN
Compressor is On.
Cylinder has finished
filling (full) -remove
cylinder (FULL light is
On in >10 mins.)
or
Cylinder is not
attached properly -
check connection
(FULL light is On in
<10 mins.).
Cylinder filling stops
Cannot fill another
cylinder until
compressor is turned
Off and On again
O
2
flowing to user if
concentrator is On.