Fillport covers not shown for clarity.
6.
Grasp the cylinder/regulator assembly in the area behind the
cylinder gauge.
7.
Position the cylinder in the compressor cradle.
8.
Align the cylinder fillport with the connector fillport.
9.
Pull UP on the outer ring (sleeve) of the connector fillport while
pushing DOWN on the cylinder/regulator assembly to couple
the cylinder fillport into the connector fillport.
The cylinder is properly connected when an audible
“click” is heard.
Disconnecting the Cylinder from the Compressor
WARNING!
–
NEVER use tools of any kind to connect/disconnect
the cylinder and the compressor. Otherwise, severe
injury and/or damage may occur.
1.
Press the compressor power switch to the Off (O) position.
2.
Grasp the cylinder/regulator assembly in the area behind the
cylinder gauge.