4 Usage
4.1 Connecting/Disconnecting the Cylinder
to/from the Compressor
Connecting the Cylinder to the Compressor
WARNING!
–
NEVER use tools of any kind to connect/disconnect
the cylinder and the compressor. Otherwise, severe
injury and/or damage may occur.
–
DO NOT drop oxygen cylinders. Use two hands when
handling/transporting oxygen cylinders. Otherwise,
injury or damage may occur.
1.
Examine the cylinder gauge. If the cylinder pressure is less than
1500 psig (10342 kPa), proceed to STEP 2.
If the cylinder pressure is greater than 1500 psig (10342
kPa), DO NOT attempt to top off this cylinder. It may
not fill.
2.
Set the conserver flow dial on the cylinder to “O”.
3.
Slowly open the shutoff valve fully (counterclockwise).
4.
Remove the cylinder and connector fillport covers.
CAUTION!
–
DO NOT connect the outlet (cannula) connection
to the HomeFill unit, otherwise the conserving
cylinder will not work properly.
5.
Momentarily push DOWN on the outer ring (sleeve) of the
connector fillport until GREEN dots are visible to reset the
connector.
If the outer ring (sleeve) is in the UP position (GREEN
dots not visible), the connector fillport will not be
able to accept the cylinder fillport. Pushing DOWN
momentarily will reset the connector fillport (GREEN
dots visible) to accept the cylinder fillport.
1176148-A
11
Содержание HomeFill HF2ECE6
Страница 10: ...HomeFill System 10 1176148 A...
Страница 12: ...HomeFill System 12 1176148 A...
Страница 31: ...Montage 1176148 A 31...
Страница 33: ...Utilisation 1176148 A 33...
Страница 44: ...Notes...
Страница 54: ...HomeFill System 54 1176148 A...
Страница 56: ...HomeFill System 56 1176148 A...
Страница 68: ...Notes...
Страница 69: ...Notes...
Страница 70: ...Notes...