Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator
5 Maintenance
5.1
Maintenance
WARNING!
Risk of Injury or Damage
Lack of product maintenance can lead to injury
and loss of product function or damage.
– Maintenance MUST be performed ONLY by
qualified personnel.
– Regular maintenance is necessary to assure
proper operation.
WARNING!
Risk of injury or damage
– The cylinder assembly MUST be disconnected
from the HomeFill and emptied of gas prior to
cleaning or disinfecting. Refer to the Emptying
the Cylinder Assembly section. The fillport
cover MUST also be installed.
– All cylinders MUST be inspected before
attempting to fill; otherwise, injury or damage
may occur. Refer to the Cylinder Prefill
Inspection section.
DANGER!
Risk of Fire
Dirt or contaminants on the cylinder assembly or
regulator can cause fire.
– DO NOT use or fill cylinder assembly if the
cylinder fillport is dirty or contaminated.
Contact your provider for service or repair.
– DO NOT clean with aromatic hydrocarbons,
including, but not limited to benzene and
toluene.
– DO NOT use cleaning/disinfecting solutions on
the fillport fitting.
– DO NOT immerse product in any kind of liquid.
– DO NOT allow any kind of liquid to get inside
openings of the regulator.
– DO NOT spray cleaning/disinfecting solutions
on product.
– All repairs MUST be performed by Invacare
Corporation or in an Invacare approved repair
center.
The cannula does not require maintenance and needs to be
replaced at the manufacturer’s recommended interval.
The aluminum cylinder must undergo hydrostatic testing
every 10 years. Refer to The Hydrostatic Testing Date section.