Invacare® EC-Track®
3.10.3 Installing and connecting control button
and power supply
Before connection of the control button and power supply,
it is necessary to dismount the covers on the motor box.
The covers are dismounted by loosening the two screws,
which holds each cover and pulling the covers horizontally
to the side.
The control switch is mounted in the room and at the
height, the user prefers. The control switch is opened with
tools, and the back side is mounted on the wall using two
screws Ø3.5x30 mm.
The wire is shortened to a suitable length and connected
to the motor box. See drawing pos.
A
.
The power supply is mounted on the wall, visible or
hidden, depending on what is most suitable. The power
supply is mounted on a Ø3.5x30 mm screw on the wall,
using the “key” hole on the back side of the power supply.
At the bottom of the power supply there is hole for a
Ø3.5x30 mm screw which makes it possible to lock the
power supply on the wall. The wire is shortened to a
suitable length and connected to the motor box. See
drawing pos.
B
.
The cover is re-attached, when the wires are connected
correctly and the control switch is tested. Remember to
mount the wires in the end-plate, in order to avoid that
the wire conflicts with the cover when it is re-attached.
See drawing pos.
C
.
CAUTION!
It is very important that there are no loose
wires hanging from the coupling.
3.10.4 Installing a traverse system with lock
fixture for transit coupling
In general a traverse system is installed as shown in
chapter
3.8 Mounting traverse system, page 21
. We
recommend that the parallel rail of the traverse system is
mounted in the ceiling in order to make the function of
the transit coupling as stable as possible.
When installing the parallel rail, remember always to place
a ceiling bracket/pendle in front of the transit coupling,
especially on the rail closest to the transit coupling. See
chapter
3.10.6 Mounting the positioning fixtures on parallel
rail S-profile and M/L-profiles, page 24
.
When the parallel rail is mounted, it is important to ensure
a correct distance to the coupling. See image in chapter
Measure sketch for transit coupling, page 23
The traverse rail is adjusted to the specific order when
it is delivered. If however it is necessary to shorten the
rail further, it must be shortened in the end opposite to
where the coupling is situated.
It is possible to use the following traverse rails together
with a transit coupling system:
•
S, M and L- profiles, under-hung
•
M and L- profiles partly between-hung
It is very important to ensure the correct free overhang
for the traverse rail spacing between the pendles and free
span from the rails to ensure that the rails can couple
correctly. See drawing above.
When the traverse system and the coupling are installed
it is very important to control the height between these
and adjust if necessary.
The lower edge of the traverse rail must not differ more
than 0.5 mm from the coupling rail. If there is a difference
in height, the coupling track must be the lowest.
It may be necessary to adjust the height differently
according to the user weight or the condition of the
ceiling, to optimize the function of the coupling.
When the lock fixtures are
installed, it is very important
that they are in line with
the end of the traverse rail
and that they are turned
correctly compared to the
fixture on the coupling,
before they are locked. See
drawing.
Lock fixtures for M- and L-profiles are delivered with a
little plate, which is important to install.
3.10.5 Installation of a single rail for a coupling
system
When the coupling is installed with a single rail in one of
the rooms, it is prepared for this, without the lock fixture,
when delivered.
When the coupling is installed with an S-profile rail, the
rail is coupled as shown in chapter
.
The pendle is mounted in the coupling fixture.
The rest of the single rail system is installed as shown in
chapters
3.2 Ceiling Mounting System, page 14
and
3.10.6 Mounting the positioning fixtures on
parallel rail S-profile and M/L-profiles
CAUTION!
It is very important that everything is adjusted
correctly and carefully fastened due to safety
and functionality.
– Test the system thoroughly after installation.