7 USAGE
Part No 1153410
39
Invacare ®CS Series Beds
7.2
Pendant Storage and Location
Ꮨ
DANGER
Electrical SHOCK Hazard.
Before working on any electrical part, such as the 9-pin pendant coupling, ensure the main power to the bed is
disconnected and the capacitor inside the control box is fully discharged.
Full capacitor discharge can be achieved by sequentially depressing the ‘Up’ and ‘Down’ buttons on the hand pendant at
least 20 times after main electrical power to the control box is cut. Ensure the hand pendant is not locked-out when
discharging it.
1.
Place the pendant holster between the top
rail bars.
2.
Slide the pendant into the holster.
3.
Alternatively, use the pendant clip to
secure the pendant to bed linen.