8.6 Electromagnetic compliance (EMC) - Guidance and manufacturer´s declaration
Guidance and manufacturer´s declaration – electromagnetic emission
This product is intended for use in the electromagnetic environment as specified below. The user of the product should
assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group I
RF emissions
CISPR 11
Class B
The product uses RF energy only for its internal function. Therefore, its RF
emissions is very low and not likely to cause any interference in nearby electronic
equipment. The product is suitable for use in all establishments, including
domestic and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Not applicable
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
Not applicable
Not applicable
Guidance and manufacturer´s declaration – electromagnetic immunity
This product is intended for use in the electromagnetic environment specified below. The user of the product should assure
that it is used in such an environment.
Immunity test
IEC 60601-1-2 test level
Compliance level
Electromagnetic environment — guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete, ceramic tile or
synthetic material.
Electrostatic
transient /
burst
IEC 61000-4-4
± 2 kV on AC power port
± 1 kV for signal and
inter-
connection
cables
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Surge
IEC 61000-4-5
± 1 kV differential mode
on AC power port
± 1 kV differential mode
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % U
T
(>100 % dip in
U
T
) for 10ms
40 % U
T
(60% dip in U
T
)
for 100 ms
70 % U
T
(30% dip in U
T
)
for 500 ms
0 % U
T
(>100 % dip in
U
T
) for 5 sec
0 % U
T
(>100 % dip in
U
T
) for 10ms
40 % U
T
(60% dip in U
T
)for 100 ms
70 % U
T
(30% dip in U
T
)
for 500 ms
0 % U
T
(>100 % dip in
U
T
) for 5 sec
Mains power quality should be that of a typical
commercial, domestic or hospital environment.
U
T
is the AC mains voltage prior to
application of the test level.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial, domestic or hospital
environment.
1532567-G
19