![Interton IBW1-V Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/interton/ibw1-v/ibw1-v_technical-manual_2080818019.webp)
Doc 0191600 rev G
Page 19 of 20
Not subject to issue control when printed
Test
Attach the CS44 prog.cable to
the USB terminal of the ear-
phone prog cable
Insert the USB terminal of the
earphone prog cable to the USB
port of the HI and turn on the HI
Insert the simulated battery into
the HI and place the HI in the test
chamber
Slide the coupler with the adap-
tor in the fixture and tighten the
bolt screw to keep it in place
(turn clockwise)
Bolt
screw
Prepare the other end of the programming cable plugs. There are two plugs, one with a “PG” mark
and the other with no mark. Connect the “ no mark” plug to the earphone when starting test
Note:
The prog cable and USB earphone prog cable are needed for testing IBW2 & 3 only. For test of IBW1, an earphone cable is
needed instead