CompactGC manual
21
Load default method
During a run, the different parameters can be reset to their start
(method) values. This is convenient when a lot of parameters
have changed, and the original settings are needed during the
run.
Reset sequence counter
See the explanation of the sequence column.
End of this run
This command stops the present run of the CompactGC. In
case of digital data-collection by EZChrom
®
, the data-
acquisition is stopped also. If the number of runs is 2 or more,
immediately the next run is started. Since the CompactGC
starts his next run immediately, and EZChrom
®
needs time to
process the acquired data, the EZChrom
®
runtime should
always be shorter than the CompactGC runtime. The time
EZChrom
®
needs between runs is dependent on the length of
the run, PC processor speed, network connections, etc.
Normally 5-10 seconds are appropriate.
Channel
If an action can be applied to more than device (temperature, flow, valves), a
channel number has to be entered.
Channel name
If name of the selected device is shown in this column.
Value
In case of a valve switching command, ‘1’ needs to be entered in this field to
activate the valve, and with value ‘0’ the valve returns to his standby position. In
combination with commands ‘T set’ and ‘FP set’ the new settings are entered
here. Commands like ‘autozero’ and ‘start chromatogram’ do not have a value.
Comment
In order to increase the readability of the RTT, comments can be entered for each
command line
For a better overview of the RTT, the operator can disable the automatic sort (on
retention time) of the table.
On the end of this page, the number of runs is entered. This number will be
executed after the start-button is pressed. A number of 999 is equal to infinite
analysis.