SULFURYL FLUORIDE MANUAL REV C
PAGE 6 OF 23
04/07/11
1]
Ready L.E.D. - illuminates when furnace reaches proper preset operating temperature.
NOTE: The “Ready” L.E.D. is not an indicator of Sensor stabilization.
2]
LO BAT. – illuminates when battery pack or AC power supply is at or below 19.0 V DC.
3]
Meter – Analog display of engineering units, parts per million (ppm).
4]
Zero – sets the zero on the meter at beginning of analysis to compensate for sensor
background current.
5]
CAL. - Sets meter to read correct engineering units when exposed to calibration gas of
certified value.
6]
ON L.E.D. – illuminates when power is turned on by function switch.
7]
FUNCTION SWITCH – selects operation and meter display:
ON – analyzer is powered and meter displays engineering units,
BAT TEST - analyzer is powered and meter displays battery pack level.
Note: Meter level must be at or above LO BAT mark for proper fuction.
8]
POWER CONNECTOR – input connector for Battery Pack or AC power supply.