Now turn on the LCT

 by pressing and holding the 

ON/OFF key (

 ) for two seconds (a). 

You will see the title 

screen (b). After about six seconds, the following screen appears (c).

a

b

c

For inital calibration, press simultaneously the two right keys for just a second (a) 

until you see the "Calibrating..." screen - which will display for three seconds (b).

a

b

Choose type of Load Cell

In order to save battery, the backlit LCD  display turns off if nothing is done for 15 seconds. To turn the dis-
play on, just push any arrow key that won't affect your settings or your current menu. To shut off the unit, 
press the Enter key (

) for a few seconds, or leave it on. It will turn off automatically after a few minutes.

Содержание LCT-1 Ultimate

Страница 1: ...LCT 1 Ultimate Load Cell Tester Issue 15 256 Revision A User s Guide www interfaceforce com...

Страница 2: ...lution and 0 1 accuracy Gain adjustment 0 1 to 5 mv V in steps of 0 01 of accuracy High resistance test 10Vdc 16 bit A D conversion Bridge test 1 25Vdc Industrial 8 pin screw connector Bridge resistan...

Страница 3: ...Email contact interfaceforce com ORDER TOLL FREE 800 947 5598 3 Follow the pin assignment and connect the Load Cell Cable to the connector In case you wish to check the isolation between the L C body...

Страница 4: ...ds the following screen appears c a b c For inital calibration press simultaneously the two right keys for just a second a until you see the Calibrating screen which will display for three seconds b a...

Страница 5: ...ne 480 948 5555 Fax 480 948 1924 www interfaceforce com Email contact interfaceforce com ORDER TOLL FREE 800 947 5598 5 Choose type of Load Cell by pressing the up or down arrows Next press the Enter...

Страница 6: ...DER TOLL FREE 800 947 5598 6 Choose L C gain by pressing the up or down arrows changing the number by increments of hundreths or by speeding through tenths if you hold the button down Note L C details...

Страница 7: ...5555 Fax 480 948 1924 www interfaceforce com Email contact interfaceforce com ORDER TOLL FREE 800 947 5598 7 Press the Enter key Initiating the LCT test Press the Enter key again for LCT LED is blink...

Страница 8: ...croll results Bridge Input and Output Resistance Results In case of a suspicious reading the LED turns red As you go through press the up key whenever you want in order to scroll to the next result un...

Страница 9: ...case of low insulation resistance Insulation resistance testing Shield to Bridge Insulation resistance testing Body to Bridge Insulation resistance testing Shield to Body End of results To scroll aga...

Страница 10: ...98 10 Applying different weights to check the linearity Model LCT 1 Ultimate is being calibrated with Model SM 100 shown above and the max weight is set for 100lbf 45 400gf For Model SM 100 the maximu...

Страница 11: ...FREE 800 947 5598 11 Applying different weights to check the linearity Apply 100g onto 45 4kg L C calculating the reading to b e 0 47 100g 45400g x 100 0 22 0 25 0 22 0 47 Apply 300g onto 45 4kg L C...

Страница 12: ...rranty repair The sender is responsible for shipping charges freight insurance and proper packaging to prevent breakage in transit The warranty does not apply to defects resulting from action of the b...