
D-0112389-I – 2022/04
Sera™ - Instructions for Use - EN
Page
63
6
Maintenance
6.1 General maintenance procedures
The performance and safety of the instrument will be maintained if the following recommendations for care
and maintenance are observed:
1. It is recommended that the instrument go through at least one annual service, to ensure that the
acoustical, electrical, and mechanical properties are correct. This should be carried out by an
authorized repairer to guarantee proper service and repair.
2. Observe that no damage is present to the insulation of the mains cable or the connectors and that it
is not exposed to any kind of mechanical load that could involve damage.
3. To ensure that the reliability of the instrument is maintained, we recommend that the operator at
short intervals, for instance once a day, performs a test on a person with known data. This person
could be the operator. For OAE testing, a daily probe integrity test is recommended to confirm the
probe is functioning properly before testing on patients.
4. If the surface of the instrument or parts of it is contaminated, it can be cleaned using a soft cloth
moistened with a mild solution of water and detergent or similar. Always disconnect the mains power
adaptor during the cleaning process and be careful that no fluid enters the inside of the instrument or
accessories.
5. After each patient examination, ensure that there has been no contamination to the parts touching
the patient. General precautions must be observed to avoid cross-contamination of disease from one
patient to another. Water should be used for frequent cleaning, but in the case of severe
contamination it may be necessary to use a disinfectant.
Please refer to the Sera
TM
Additional Information for further information about cleaning and
maintenance of the equipment and accessories.
6.2 General maintenance procedures
•
Before cleaning always switch off and disconnect from the power supply
•
Use a soft cloth lightly dampened with cleaning solution to clean all exposed surfaces
•
Do not allow liquid to meet the metal parts inside the earphones/headphones
•
Do not autoclave, sterilize, or immerse the instrument or accessory in any fluid
•
Do not use hard or pointed objects to clean any part of the instrument or accessory
•
Do not let parts that have been in contact with fluids dry before cleaning
•
Rubber ear-tips or foam ear-tips are single use components
Recommended cleaning and disinfection solutions:
•
Warm water with mild, non-abrasive cleaning solution (soap)
Содержание Sera
Страница 1: ...Instructions for Use EN Sera Science made smarter 8501165 D 0112389 I 2022 04...
Страница 2: ......
Страница 6: ......
Страница 14: ...D 0112389 I 2022 04 Sera Instructions for Use EN Page 8...