3.10 Trouble shooting
LCD Display or Problem
Observed
Cause
Solution
No Water in Tank
External water tank empty
Refill water tank with water at less
than 75°F (24°C). Check hose
connections. Select Cool / Warm to
retry operation
Fill Cool Water
Water in external tank is warmer
than the desired irrigation
temperature
Add cooler water to the external
water tank then select Cool / Warm
to retry operation
Irrig. Timeout
Irrigator left on for 10 minutes
without performing another
irrigation, will return to Standby
screen
Press Cool or Warm to initiate
irrigation procedure, otherwise none
Select Cool / Warm to retry
operation
No water comes out of the
irrigator handle after
irrigator displays “Ready”
and pressing the irrigation
button.
Possible blockage in the irrigator
handle.
Disconnect the handle from the
irrigator. Push air with a syringe into
the blue tube to flush out water and
impurities. Discard water coming out
of the red tube.
Water dribbles out of the
irrigation handle when the
irrigator is running in Cool or
Warm mode. This happens
before “Ready” state is
reached.
Fault in the electronics power
module for the handle.
Check the handle’s electrical
connector on the back panel is
properly connected.
Too Warm
Residual warm water in the
irrigator preventing reaching the
desired temperature
Stop current operation by pressing
Cool / Warm. Then retry desired
temperature Cool / Warm. Aqua Stim
will purge the internal water by filling
from the external water tank.
Handle does not light or
respond to button press.
Electrical connector not making
good contact.
Power-down irrigator then check
connections on back.
Call TechSupport
(note additional error
message)
Internal problem requiring
technical support for assistance
Call Interacoustics Technical
Support with the
additional error
message
on the display (e. g.
“Level Sense Error” etc).
Prior to calling Interacoustics Technical Support note the firmware version of the Aqua Stim irrigator. This
is shown on the display briefly when the Aqua Stim first starts.