13

 

 

4.3  Daily 

External surface of the device which is generally used by the healthcare professional is to be cleaned 

daily with recommended cleaning solution (refer section 4. 1. 2).  

 

Cleaning procedure:

 Wipe off the external surface with a disposable, clean, non-linting cloth which is 

damped in the cleaning solution until all visible soil is removed. Ensure that moisture doen’t enter the 

critical areas of device. 

Cleaning solution should be changed at each cleaning session and when visibly 

soiled. 

 

 

 

4.4  Quarterly maintenance 

Drain the water completely from the irrigator. Remove the lid from the water tank and use a paper towel to 

reach inside and wipe down the inside wall of the water reservoir which is used for heat management. If 

the paper towel does not show a biofilm, then refill the reservoir with 

distilled or demineralized water

 for 

routine usage of irrigators 
If the biofilm is observed, fill the water reservoir of irrigator with the disinfectant 

(70-85% v/v Isopropyl 

alcohol)

. Power on the irrigator for 30 minutes with cool Irrigation. The water pump will circulate the 

disinfectant through the lines as long as power is on. After 30 minutes, drain the solution from the 

irrigator. Then, fill the irrigator with 

distilled or demineralized water 

and run couple of cool irrigation 

cycles. Drain the irrigator to ensure that the disinfectant is washed out. Then refill the reservoir with 

distilled or demineralized water 

for routine usage of irrigator.  

 
 

4.5  Annual maintenance  

The Annual Cleaning Procedure should be performed by a qualified service technician.  

 

4.5.1  Air filter cleaning 

Check the exhaust fan filter for debris or dust. The filter can be removed and cleaned after turning off 

power to the irrigator and detaching the plastic retaining bracket on the exhaust fan.  
Check the air inlet filter for debris or dust. Remove the grommet and tubing using a pair of needle-nosed 

pliers. Blow out debris with a can of compressed air.  
 

4.5.2  Water filter cleaning 

The Air Fx caloric irrigator uses an inline water filter to collect debris. The water filter cone should be 

replaced once a year by a qualified service technician. Refer to the section 4. 6 Replacing the Water 

Filter.  

 

 

4.5.3  Irrigator cleaning 

Annually the Air Fx should be cleaned to remove any biofilm and mineral deposits. The Annual Cleaning 

Procedure should be performed by a qualified service technician.  

 

 

4.6  Annual irrigator verification 

The Air Fx flow rate and air temperature should be verified by a qualified service technician. If the irrigator 

is outside of specifications for flow rate or temperature, then the irrigator must be sent back to 

Interacoustics for calibration.  

 

  

 

 

Содержание Air Fx

Страница 1: ...Instructions for Use US Air Fx Caloric Irrigator Science made smarter 8525533 D 0128226 A 2020 12...

Страница 2: ...thout notice No part of this document may be reproduced or transmitted in any form or by any means without a prior written permission of Interacoustics A S FireWire is a registered trademark of Apple...

Страница 3: ...re 9 3 6 Perform the irrigation 9 3 7 Safety 10 3 8 Turn off the irrigator 10 3 9 Draining the water reservoir 10 4 MAINTENANCE 12 4 1 General cleaning procedure 12 4 1 1 General Precautions 12 4 1 2...

Страница 4: ...the caloric test with the correct irrigation temperature All personnel who operate the Air Fx should familiarize themselves with the contents of this manual prior to using the Air Fx with a patient A...

Страница 5: ...C 10 Amperes Speculum 2 75 mm Speculum 4 25 mm 60cc syringe and tube USB Cable 3m 9 8ft User s Manual Drain Fill Kit Hose Management Tape 1 4 Warnings and precautions Throughout this manual the follow...

Страница 6: ...he invoice serial number and a detailed report of the issue For any on site service related information please contact your local distributor If the system components are to be returned for service pl...

Страница 7: ...sed at close distances to the equipment Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Performance Conditions...

Страница 8: ...ons for Use Observe precautions for handling electrostatic sensitive devices It is against the law to dispose of electronic devices in the trash The crossed out wheelie bin symbol shows that the compo...

Страница 9: ...ol Irrigation to be performed returns to standby state C Adjust temperature set point up 1 C D Adjust temperature set point down 1 C E Hose connection F Handle support G LCD output screen Figure 2 Rea...

Страница 10: ...e tube to the Fill Drain Water Outlet Fully push the tube into the Fill Drain Water push in connector about 1 2cm of tubing Gently pull on the tubing to confirm the tube is secure 5 Switch the Fill Dr...

Страница 11: ...urn on the irrigator Turn on the power switch on the back panel The Air Fx will initialize show the firmware version and then go into the standby state Select Cool Warm Air Fx v1 4 3 4 Attaching the s...

Страница 12: ...mperature e g 23 4 C Cool 24 C 23 4 C Irrigation will be available after the desired temperature has been reached and stabilized Once this occurs the Air Fx will emit an indicator tone and will displa...

Страница 13: ...n the irrigator and isolate the problem ambient temperature too low bubbles in the water line distilled or demineralized water level too low etc Contact Interacoustics Service Department or the local...

Страница 14: ...ube Squirt any distilled or demineralized water into a waste container Reconnect the syringe to the tube Repeat this process until all of the distilled or demineralized water is removed from the inter...

Страница 15: ...ion control The WHO s recommended level of decontamination for non critical devices is cleaning Disinfection and sterilization are not recommended However in case of an epidemic outbreak disinfection...

Страница 16: ...r Then fill the irrigator with distilled or demineralized water and run couple of cool irrigation cycles Drain the irrigator to ensure that the disinfectant is washed out Then refill the reservoir wit...

Страница 17: ...ed Interacoustics service representative altered in any way so as in Interacoustics judgement to affect its stability or reliability subject to misuse or negligence or accident or which has had the se...

Страница 18: ...ctric and electronic waste as unsorted municipal waste Electric and electronic waste may contain hazardous substances and therefore has to be collected separately Such products will be marked with the...

Страница 19: ...D 0128226 A 2020 12 Air Fx Caloric Irrigator Instructions for Use US Page 16...

Страница 20: ...faster 3 m 9 8 ft cable Case dimensions 35W x 32D x 22H cm 13 8 x 12 6 x 8 7 in Air hose length 3 m 9 8 ft Weight 8 kg 18 lb Voltage 110 130 VAC 50 60 Hz 220 240 VAC 50 60 Hz Fuses 2 each 110VAC unit...

Страница 21: ...ons or decreased electromagnetic immunity of this equipment and result in improper operation The list of accessories and cables can be found in this section Portable RF communications equipment includ...

Страница 22: ...omagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Air Fx can help prevent electromagnetic interferences by maintaining a minimum distance between p...

Страница 23: ...nvironment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or residential environment Vol...

Страница 24: ...meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the...

Страница 25: ...s or fax No to whom Interacoustics may confirm reception of the returned goods The above mentioned item is reported to be dangerous to patient or user 1 In order to ensure instant and effective treatm...

Отзывы: