2.5  Preparing the irrigator for first use 

The Air Fx is ventilated by a fan at the back of the device. Do not place the Air Fx near a radiator or other 

heating source. Provide at least 10 cm (4 in) of free space behind the unit to provide adequate circulation.  

 

Filling the water reservoir 

The Air Fx will be shipped from Interacoustics without any 

distilled or demineralized water

 in the reservoir. 

Please fill according to the following instructions.  

 

 

The Air Fx cannot be used without adequate 

distilled or demineralized water

 

present in the water reservoir. The 

distilled or demineralized water 

must be 

added before applying power. Failure to maintain the proper level of fresh 

distilled or demineralized water

 can result in permanent damage to the unit. 

 

 

1.  Check that the irrigator is powered OFF.  

2.  Remove the cap to the water reservoir on the top of the irrigator. Do not force the cap or use mechanical 

tools to remove the cap.  

3.  Fill the reservoir with distilled water up to 1cm from the bottom of the cap. Take care not to spill any 

water into the unit except into the water reservoir to prevent damage to the circuitry and prevent any 

electrical shock to the user. If any water is spilled, wipe it up immediately.  

4.  Purge air from the hoses, pump and handle by connecting the included 60cc syringe & tube to the Fill / 

Drain Water Outlet. Fully push the tube into the Fill / Drain Water push-in connector (about 1” (2cm) of 

tubing). Gently pull on the tubing to confirm the tube is secure.  

5.  Switch the Fill / Drain Valve to the Fill / Drain position.  

6.  Draw back the syringe to draw the air out of the internal tubing and draw distilled water into the syringe. 

If there is a train of air bubbles or there is no suction, then set the Fill / Drain Valve to the Run position 

and reconnect the tube. Once reconnected, set the Fill / Drain Valve to the Fill / Drain position.  

7.  Disconnect the syringe from the tube and keep the tube held above the irrigator to prevent the water 

from exiting the tube. Squirt any distilled water drawn back into the reservoir. Reconnect the syringe to 

the tube.  

8.  Repeat steps 6 to 7 two more times to draw the remaining air out of the lines.  

9.  Set the Fill / Drain Valve to the Run position.  

10. Disconnect the syringe. The tube can be removed by pushing on the sides of the tubing connector’s grey 

collar around the tubing while simultaneously pulling on the tubing.  

11. Place the cap back on the water reservoir.  

12. Attach the AC power cord to the back of the irrigator and switch on the Air Fx using the power switch 

next to the power cord. Water will automatically be pumped through the hoses into the handle and 

returned to the reservoir.  

13. Test system by running irrigator on the warm cycle then the cool cycle. If the irrigator is able to do this, 

then it is ready to use. If the irrigator is unable to reach the desired temperature, then repeat the fill 

instructions.  

 

 

Содержание Air Fx

Страница 1: ...Instructions for Use US Air Fx Caloric Irrigator Science made smarter 8525533 D 0128226 A 2020 12...

Страница 2: ...thout notice No part of this document may be reproduced or transmitted in any form or by any means without a prior written permission of Interacoustics A S FireWire is a registered trademark of Apple...

Страница 3: ...re 9 3 6 Perform the irrigation 9 3 7 Safety 10 3 8 Turn off the irrigator 10 3 9 Draining the water reservoir 10 4 MAINTENANCE 12 4 1 General cleaning procedure 12 4 1 1 General Precautions 12 4 1 2...

Страница 4: ...the caloric test with the correct irrigation temperature All personnel who operate the Air Fx should familiarize themselves with the contents of this manual prior to using the Air Fx with a patient A...

Страница 5: ...C 10 Amperes Speculum 2 75 mm Speculum 4 25 mm 60cc syringe and tube USB Cable 3m 9 8ft User s Manual Drain Fill Kit Hose Management Tape 1 4 Warnings and precautions Throughout this manual the follow...

Страница 6: ...he invoice serial number and a detailed report of the issue For any on site service related information please contact your local distributor If the system components are to be returned for service pl...

Страница 7: ...sed at close distances to the equipment Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Performance Conditions...

Страница 8: ...ons for Use Observe precautions for handling electrostatic sensitive devices It is against the law to dispose of electronic devices in the trash The crossed out wheelie bin symbol shows that the compo...

Страница 9: ...ol Irrigation to be performed returns to standby state C Adjust temperature set point up 1 C D Adjust temperature set point down 1 C E Hose connection F Handle support G LCD output screen Figure 2 Rea...

Страница 10: ...e tube to the Fill Drain Water Outlet Fully push the tube into the Fill Drain Water push in connector about 1 2cm of tubing Gently pull on the tubing to confirm the tube is secure 5 Switch the Fill Dr...

Страница 11: ...urn on the irrigator Turn on the power switch on the back panel The Air Fx will initialize show the firmware version and then go into the standby state Select Cool Warm Air Fx v1 4 3 4 Attaching the s...

Страница 12: ...mperature e g 23 4 C Cool 24 C 23 4 C Irrigation will be available after the desired temperature has been reached and stabilized Once this occurs the Air Fx will emit an indicator tone and will displa...

Страница 13: ...n the irrigator and isolate the problem ambient temperature too low bubbles in the water line distilled or demineralized water level too low etc Contact Interacoustics Service Department or the local...

Страница 14: ...ube Squirt any distilled or demineralized water into a waste container Reconnect the syringe to the tube Repeat this process until all of the distilled or demineralized water is removed from the inter...

Страница 15: ...ion control The WHO s recommended level of decontamination for non critical devices is cleaning Disinfection and sterilization are not recommended However in case of an epidemic outbreak disinfection...

Страница 16: ...r Then fill the irrigator with distilled or demineralized water and run couple of cool irrigation cycles Drain the irrigator to ensure that the disinfectant is washed out Then refill the reservoir wit...

Страница 17: ...ed Interacoustics service representative altered in any way so as in Interacoustics judgement to affect its stability or reliability subject to misuse or negligence or accident or which has had the se...

Страница 18: ...ctric and electronic waste as unsorted municipal waste Electric and electronic waste may contain hazardous substances and therefore has to be collected separately Such products will be marked with the...

Страница 19: ...D 0128226 A 2020 12 Air Fx Caloric Irrigator Instructions for Use US Page 16...

Страница 20: ...faster 3 m 9 8 ft cable Case dimensions 35W x 32D x 22H cm 13 8 x 12 6 x 8 7 in Air hose length 3 m 9 8 ft Weight 8 kg 18 lb Voltage 110 130 VAC 50 60 Hz 220 240 VAC 50 60 Hz Fuses 2 each 110VAC unit...

Страница 21: ...ons or decreased electromagnetic immunity of this equipment and result in improper operation The list of accessories and cables can be found in this section Portable RF communications equipment includ...

Страница 22: ...omagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Air Fx can help prevent electromagnetic interferences by maintaining a minimum distance between p...

Страница 23: ...nvironment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or residential environment Vol...

Страница 24: ...meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the...

Страница 25: ...s or fax No to whom Interacoustics may confirm reception of the returned goods The above mentioned item is reported to be dangerous to patient or user 1 In order to ensure instant and effective treatm...

Отзывы: