D-0128226-A – 2020/12
Air Fx Caloric Irrigator – Instructions for Use - US
Page
11
4. Draw back the syringe to draw the
distilled or demineralized water
out of the internal tubing and
water reservoir.
5. Disconnect the syringe from the tubing and keep the tube held above the irrigator to prevent the water
from exiting the tube. Squirt any
distilled or demineralized water
into a waste container. Reconnect
the syringe to the tube. Repeat this process until all of the
distilled or demineralized water
is
removed from the internal lines, until only air is drawn into the syringe.
6. Set the Fill / Drain Valve to the Run position.
7. Disconnect the syringe and tubing. The tubing can be removed by pushing in the gray gasket around
the tubing with one’s finger nails while pulling on the tubing to remove.
8. Replace the cap back on the water reservoir.