background image

6. Calibration Instructions

The TT820D comes factory calibrated. If you need to re-calibrate the unit, two
options are available. The first one is the “ One Point Calibration ” and the second
one is the “ Advanced Calibration “.

One Point Calibration

In some cases, a single point calibration is sufficient especially when a process is at
a fixed set point.

1. With a temperature standard, verify the correct process temperature and

compare it to the TT820D reading.

2. With the “ UP ” and “ DW “ push buttons, located on the front of the

TT820D transmitter, adjust current output to the value read by the standard.

Advanced Calibration

To perform the “ Advanced Calibration ” you will need :

The MIST TT820D Programming Kit. ( ordered separatly )

•  A milliamp indicator or a loop-powered calibrator.
•  A temperature bath.

A standard reference temperature sensor.

Test leads

3

2

4-20 mA

Indicator

TT820D

Communication

Interface

AC Adapter

110V AC

RS232

1.  Connect the TT820D as per the wiring diagram shown above.

2.  Start the MIST T/C software.

3.  Go in the Programming section of the software by clicking on the 

“ Program ” button.

4.  Set the temperature bath to the minimum range of the TT820D transmitter 

( Ex.: 0

o

C = 4 mA )

5.  With the temperature standard, verify the bath temperature and enter the 

value in the “ First Point” text box located in the “ 2 Points Calibration ” 
section of the “ Programming ” window.

6.  Immerse the TT820D probe in the temperature bath and wait for the 

output stabilizes.

7.  Click on the “ Read ” button associated to the “ First Point ” calibration.

+i

-i

Tx

Rx

+

_

Tx
Rx

E
F

G
H

D

C

+TC

-TC

2.448 ”

[ 62.18 ]

2.52 ”

[ 64.00 ]

.670 ”

[ 17.02 ]

.490 ”

[ 12.45 ]

4. Dimensions

+

_

Power Supply

12-36V DC

4-20 mA

Indicator

The diagram above represents the connection method to provide the 4-20 mA with
a current loop output. The connections should be made in the following order :

1. Connect the Thermocouple to the « C » and « D » terminals on the TT820D.

Be sure to respect the polarity of the thermocouple ( C to Neg. D To Pos. ) 
otherwise the output will not correspond to the right reading.

2. Connect the positive side of the power supply ( 12-36 VDC ) to the « E » 

terminal on the TT820D.

3. Connect the positive side of the current loop indicator to the « F » terminal

on the TT820D.

4. Connect the negative side of the current loop indicator to the negative side

of the power supply ( 12-36 VDC ). 

Note :

i )  It takes about 30 seconds, due to the filtering features, before the output 

signal stabilizes to its value.

ii ) The indicator could be replaced by a controller, a data logger, etc. 

In any case, be sure to respect the maximum load allowed on the loop, 
given by :

Rmax. = [ ( Vsupply – 12V ) / 20mA ] 

5. Electrical Connections

TT820D

E

F

D

C

+TC

-TC

Отзывы: