InTemp® CX400 Series Temperature Logger Manual
19996-J
InTemp CX400 series loggers measure temperature in indoor monitoring applications. Designed
to meet World Health Organization (WHO) and Centers for Disease Control and Prevention
(CDC) guidelines, this logger is ideal for clinical applications, such as vaccine storage and
pharmaceutical manufacturing. This Bluetooth® Low Energy-enabled logger is designed for
wireless communication with a mobile device. Using the InTemp app, you can easily configure
the logger with one of four preset configurations designed for ambient storage, clinical
refrigerator, freezer, or refrigerator monitoring or you can set up a custom configuration for
other applications. You can also quickly perform daily logger checks, download reports, and
monitor tripped alarms. Use the built-in LCD screen to check the current temperature, daily
maximum or minimum temperatures, logging status, battery use, and more. You also have the
option to use InTempConnect® to track logger deployments and automatically upload logger
data to build custom reports for further analysis. The CX402 model uses a 2- or 4-meter probe
and is available with a 5, 15, or 30 mL glycol bottle (bottle holder included). It also contains an
internal sensor for ambient temperature monitoring. The CX403 model is available with an
internal sensor only.
Specifications
External Temperature Probe
Range
-40° to 100°C (-40° to 212°F)
Accuracy
CX40x-Txxx and CX402-BxM:
±1.0°C from -40° to -15°C (±1.8°F from -40° to 5°F)
±0.5°C from -15° to 80°C (±0.9°F from 5° to 176°F)
±0.7°C from 80° to 100°C (±1.3°F from 176° to 212°F)
CX40x-VFCxxx:
±1.0°C from -40° to -22°C (±1.8°F from -40° to -8°F)
±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F)
±1.0°C from 50° to 100°C (±1.8°F from 122° to 212°F)
Resolution
0.024°C at 25°C (0.04°F at 77°F)
Drift
<0.1°C (0.18°F) per year
Response Time
Probe only: 1 minute 30 seconds typical to 90% in air moving 1 m/s
With 5 mL bottle: 10 minutes typical to 90% in air moving 1 m/s
With 15 mL bottle: 20 minutes typical to 90% in air moving 1 m/s
With 30 mL bottle: 25 minutes typical to 90% in air moving 1 m/s
NIST Calibration
CX40x-Txx and CX402-BxM: Single point NIST calibration, probe and
logger body
CX40x-VFCxxx: Single point NIST calibration, probe only
Cable Length
2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable
Probe Dimensions
Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches)
long, 3.18 mm (0.125 inches) diameter
Ambient Temperature Sensor
Range
-30° to 70°C (-22° to 158°F)
Accuracy
CX40x-Txxx and CX402-BxM:
±0.5°C from -15° to 70°C (±0.9°F from 5° to 158°F)
±1.0°C from -30° to -15°C (±1.8°F from -22° to 5°F)
CX40x-VFCxxx:
±1.0°C from -30° to -22°C (±1.8°F from -22° to -8°F)
±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F)
±1.0°C from 50° to 70°C (±1.8°F from 122° to 158°F)
Resolution
0.024°C at 25°C (0.04°F at 77°F)
Drift
<0.1°C (0.18°F) per year
Response Time
15 minutes typical to 90% in air moving 1 m/s
Logger
Radio Power
1 mW (0 dBm)
Transmission Range
Approximately 30.5 m (100 ft) line-of-sight
Wireless Data Standard
Bluetooth Low Energy (Bluetooth Smart)
InTemp CX400 Temperature
Loggers
Models:
•
CX402-T205 & CX402-VFC205,
2 meter probe and 5 mL glycol
bottle
•
CX402-T215 & CX402-VFC215,
2 meter probe and 15 mL
glycol bottle
•
CX402-T230 & CX402-VFC230,
2 meter probe and 30 mL
glycol bottle
•
CX402-T405 & CX402-VFC405,
4 meter probe and 5 mL glycol
bottle
•
CX402-T415 & CX402-VFC415,
4 meter probe and 15 mL
glycol bottle
•
CX402-T430 & CX402-VFC430
4 meter probe and 30 mL
glycol bottle
•
CX402-T2M, CX402-B2M, &
CX402-VFC2M, 2 meter probe
•
CX402-T4M, CX402-B4M, &
CX402-VFC4M, 4 meter probe
•
CX403
Included Items:
•
Double-sided tape (for models
with glycol bottles)
•
Two AAA 1.5 V alkaline
batteries
•
Battery door and screw
•
NIST Certificate of Calibration
Required Items:
•
InTemp app
•
Device with iOS or Android™
and Bluetooth