![Intellisystem ThermalTronix TT-1260S-UTCM Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/intellisystem/thermaltronix-tt-1260s-utcm/thermaltronix-tt-1260s-utcm_user-manual_2074817011.webp)
10
X: XXX
Dead pixel position in X direction;
Y: XXX
Dead pixel position in Y direction;
Ent:
Save dead pixel correction parameter and exit the menu;
Esc:
Exit dead pixel correction menu without saving change.
Note:
Move the crosshair to the dead pixel by pressing
UP/DOWN
buttons and then press
C
button to replace the dead pixel.
3.1.5 Factory default menu
User can restore the factory default setting via this menu. Press
F
and
DOWN
buttons to activate
the factory default menu.
Def: N Revise: N
AutoDoZero: N
Figure.3.1.5.1 Factory default menu interface
Def: N/Y
Display option of restoring factory default, such as dead pixels and
language.
Revise: N/Y
Enable or disable the revise function.
AutoDoZero: N/Y
Enable or disable the function of zeroing automatically.
Note
: use
UP/DOWN
buttons to choose option value. Keep pressing
MENU
button until the
menu disappeared to exit.