3. INITIAL SETUP
3.1 INTERNAL FUNCTIONS AND SETTING METHODS
INTERNAL FUNCTION TABLE
Function Symbol
Description
1 F1
Span
value
reading
2
F2
Full display segment check
3 F3
Scale
configuration
4
F4
Auto power off setting
9
F9
Auto tare setting
19
F19
1/30000 Verification Mode**
20
F20
Gravity Compensation-place of calibration**
21
F21
Gravity Compensation-place to be used**
HOW TO ENTER THE REQUIRED FUNCTION MODE
a. Turn scale off.
b. Press and hold TARE, then turn scale on. Scale display F.1
c. Press TARE until the required function number appears.
d. Press MODE
e. Press MODE until the required setting appears.
f. Press TARE to confirm.
g. Repeat step c to f for other function setting, or
h. Press ON/ZERO to save settings and return to normal operation.
F1 Span Value Reading
a. Simply enter F1 to read the A/D counts.
b. Press
ON/ZERO to clear the A/D counts, apply test mass onto platter,
the span value of test mass will be displayed.
F2 Full Display Segment Check
When function is entered, scale will firstly display the software
version, and then displays all segments later.
Check and make sure that no segments are missed.
F3 Scale Configuration
a. Enter F3, scale displays the offset value.
b. Press and hold MODE, available scale capacity and min.
division
appears.
SELECT DISPLAY RESOLUTION
c. Press and hold MODE to change the display resolution.
SELECT WEIGHT UNITS
d. when desired display resolution is selected, press MODE
Содержание AXM-1500
Страница 5: ...5 2 3 GENERAL SPECIFICATION 2 3 1 Overall View...
Страница 15: ...15 5 ELECTRICAL CIRCUITRY 5 1 SCHEMATICS...
Страница 16: ...16...
Страница 17: ...17...
Страница 18: ...18...
Страница 19: ...19...
Страница 21: ...21 5 2 PCB LAYOUT AXM 10 1 TOP LAYER AXM 10 1 BOTTOM LAYER AXM 10 1 TOP OVERLAY AXM 10 1 BOTTOM OVERLAY...
Страница 22: ...22 AXM 40 1 TOP LAYER AXM 40 1 BOTTOM LAYER AXM 40 1 TOP OVERLAY AXM 40 1 BOTTOM OVERLAY...
Страница 27: ...27 7 APPENDIX...