1. PUT THE WEIGHT (1/3 OF MAX RANGE) TO CHECK IN THE CENTER OF THE PLATE, AND
PRESS TARE.
2. MOVE THE WEIGHT IN THE POINT 2A, READ AND WRITE DOWN THE VALUE.
3. MOVE THE WEIGHT IN THE CENTER AND PRESS TARE.
4. MOVE THE WEIGHT IN THE POINT 2B, READ AND WRITE DOWN THE VALUE.
5. TAKE THE TOOL AND REGULATE AS ILLUSTRATED IN THE TABLE BELOW
IF THE CORNER 2A POSIT IVE(+) THEN
TURN THE TOOL CLOCKWISE
(SCREW A)
IF THE CORNER 2A NEGATIVE(-) THEN
TURN THE TOOL ANTICLOCKWISE
(SCREW A)
IF THE CORNER 2B PO) THEN
TURN THE TOOL ANTICLOCKWISE
(SCREW B)
IF THE CORNER 2B NEGATIVE(-) THEN
TURN THE TOOL CLOCKWISE
(SCREW B)
Attention: if the corner opposite can not corrected, check if the mechanical group is ok.