Active run step (orange)
Parameters are listed
in the lower half
From the Main Menu, use the
touch wheel
to scroll to your
desired program.
Press
OK
to enter the selected
program.
Insert the GRIPTIPS into the liquid to be transferred.
Press and release
RUN
to aspirate the volume
selected in the first step.
Your pipette will prompt you to press
RUN
at each
successive step.
Setting/changing parameters
Press
OK
to Edit the program. A list of editable steps
is displayed.
Select a step and press
OK
.
Use the
touch wheel
to set the
value and press
OK
.
Press
to save your settings.
Installation of second stage
The
second stage
can be used as a second position, e.g., place source
liquids on the
base stage
and slid a well plate on the
second stage
for
filling.
•
Remove two
magnetic covers
(
c
) by using a GRIPTIP.
•
Insert the
second stage
with the desired side (96 or 384 format) facing
upwards.
Two Step Blowout
During aspiration, the piston of your electronic pipette moves up (1).
During dispensing, the piston returns to the initial position (0). During the last
dispense of a program, it automatically moves further down (2) and blows the
remaining liquid from the tip (Blowout
).
When the piston returns to the initial position (0), a small amount of air is
aspirated, provided the tip is no longer immersed in the liquid (Blowin
).
Note: Manually delay the blowin by holding
RUN
pressed during the last dispense. Remove the
tips from the vessel and release
RUN
to start blowin.
Maintenance
Always switch off the device and disconnect from the electricity supply when carrying out
maintenance work.
Clean the MINI 96 housing with a lint-free cloth lightly soaked with mild soap solution in distilled
water or with a 70 % solution of isopropanol or ethanol.