129951_V02_QSG_MINI_96.docx 

3/4 

        

 

 

 
 

Active run step (orange) 

 
 

Parameters are listed  
in the lower half

 

From the Main Menu, use the  

touch wheel

 to scroll to your 

desired program.  

Press 

OK

 to enter the selected 

program.

  

Insert the GRIPTIPS into the liquid to be transferred. 
Press and release 

RUN 

to aspirate the volume 

selected in the first step.  

Your pipette will prompt you to press 

RUN 

at each 

successive step. 

 

Setting/changing parameters 

 

 

 

      

Press 

OK

 to Edit the program.  A list of editable steps  

is displayed.  
Select a step and press 

OK

.

 

Use the 

touch wheel

 to set the 

value and press 

OK

.  

Press 

 to save your settings.

 

 

Installation of second stage 

 

 

 

The 

second stage

 can be used as a second position, e.g., place source 

liquids on the 

base stage

 and slid a well plate on the 

second stage

 for 

filling. 

  Remove two 

magnetic covers

 (

c

) by using a GRIPTIP. 

  Insert the 

second  stage

  with the  desired side  (96  or  384  format)  facing 

upwards.  

 

Two Step Blowout 

 

 

During aspiration, the piston of your electronic pipette moves up (1).  

During dispensing, the piston returns to the initial position (0). During the last 
dispense of a program, it automatically moves further down (2) and blows the 
remaining liquid from the tip (Blowout

). 

When  the  piston  returns  to  the  initial  position  (0),  a  small  amount  of  air  is 
aspirated, provided the tip is no longer immersed in the liquid (Blowin

).  

Note: Manually delay the blowin by holding 

RUN 

pressed during the last dispense. Remove the 

tips from the vessel and release 

RUN 

to start blowin. 

 

Maintenance 

 

 

Always switch off the device and disconnect from the electricity supply when carrying out 
maintenance work.

 

Clean the MINI 96 housing with a lint-free cloth lightly soaked with mild soap solution in distilled 
water or with a 70 % solution of isopropanol or ethanol. 

Содержание 4801

Страница 1: ...icable regulations must be observed 1 The pipette may only be used by properly trained personnel in a manner specified by INTEGRA Biosciences 2 Do not use the instrument near flammable material or in an atmosphere with danger of explosion 3 Do not immerse the pipetting unit in liquid Avoid pipetting of liquids emitting corrosive vapors 4 Servicing work and repairs may only be performed by INTEGRA ...

Страница 2: ... height for ex the bottom of a well plate select Pipetting Height With tips attached move the pipetting unit down to the desired pipetting height Press Set to save your setting Running a pipetting program Select from predefined programs that you can adapt easily or create multi stepped custom programs Program Description Pipet Liquid transfers when aspirate and dispense volumes are equal Repeat Di...

Страница 3: ...ing Remove two magnetic covers c by using a GRIPTIP Insert the second stage with the desired side 96 or 384 format facing upwards Two Step Blowout During aspiration the piston of your electronic pipette moves up 1 During dispensing the piston returns to the initial position 0 During the last dispense of a program it automatically moves further down 2 and blows the remaining liquid from the tip Blo...

Страница 4: ...s com info integra biosciences com Direct sales country Telephone Mail China 86 21 5844 7203 info cn integra biosciences com Denmark Sweden 45 3173 5373 info nordic integra biosciences com France 33 1 34 30 76 76 info fr integra biosciences com Germany Austria 49 6409 81999 15 info de integra biosciences com Japan 81 3 5962 4936 info jp integra biosciences com Switzerland 41 81 286 95 30 info ch i...

Отзывы: