Appendix
178
Appendix
Non-pyrogenic
: Fluid pathway only
Type BF applied part
: Protection from electrical shock
Maximum infusion pressure
: 10 psi
Maximum volume infused under single fault conditions
: 0.0 U
Flow Capability
:
Basal: Programmable by the user in 0.05 U increments of up to 30.0 U per
hour
Bolus Rate: 1.5 units per minute. Dose range from 0.05 to 30.0 units
Delivery accuracy
(tested per IEC 60601-2-24):
Basal: ± 5% at rates ≥ 0.05 U/hr
Bolus: ± 5% for amounts ≥ 1.0 unit
± 0.05 units for amounts < 1.0 unit
Note:
The user should consider bolus dose accuracy when setting a bolus
dose. When using the lowest bolus dose allowable (0.05 units), the actual bolus
delivered may be as low as 0.00 units or as high as 0.10 units.
Observation Interval (min)
Err
or (%)
EpMax
EpMin
Effective Date: 27SEP2019, ECO-000240
Содержание INT2-D001-MM-OUS2
Страница 11: ...x Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Страница 17: ...xvi Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Страница 37: ...20 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Страница 53: ...36 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Страница 73: ...56 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Страница 85: ...68 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Страница 141: ...124 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Страница 177: ...160 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Страница 211: ...194 Thispageintentionallyleftblank Effective Date 27SEP2019 ECO 000240...
Страница 221: ...My Notes Effective Date 27SEP2019 ECO 000240...