49
Excessive amount of air inside nozzle
If you experience
problems drawing insulin
from your cartridge
or vial, it is likely that
the nozzle contains an
excessive amount of air.
To remove the air from
the nozzle, first remove
the adaptor with the
cartridge or vial from
the nozzle.
CAUTION!
Never release the InsuJet™ with air inside
the nozzle. This is called DRY-FIRING and
can damage the InsuJet™ beyond repair.
Keep the InsuJet™ upright. While
holding the front of the device, rotate
the back in the direction of arrow ‘1’
untill the green marking is visible in the
dosage lens.
Follow the steps described in the
section ‘PREPARING THE INSUJET
FOR ADMINISTRATION’ to draw insulin
into the nozzle.