Instruction Manual VGC301A Controller
InstruTech, Inc.
Page 23
7.2
Non-Linear Analog Output - mbar
Non-Linear analog output for selected gases - Engineering units in mbar
True Pressure
N
2
Ar
He
O
2
CO
2
KR
Freon12 Freon22
D
2
Ne
CH
4
0 mbar
0.3751
0.375
0.375
0.375
0.375
0.375
0.375
0.375
0.375
0.375
0.375
.0001 mbar
0.3759
0.3757
0.376
0.376
0.376
0.376
0.376
0.376
0.376
0.3757
0.3766
.0003 mbar
0.3768
0.376
0.377
0.377
0.377
0.377
0.378
0.378
0.377
0.3763
0.378
.0006 mbar
0.3795
0.378
0.379
0.38
0.381
0.377
0.382
0.381
0.381
0.3782
0.3825
.0013 mbar
0.384
0.381
0.382
0.384
0.385
0.379
0.388
0.388
0.386
0.381
0.3896
.0027 mbar
0.3927
0.387
0.389
0.392
0.395
0.384
0.401
0.4
0.396
0.388
0.403
.0067 mbar
0.4174
0.403
0.409
0.417
0.412
0.395
0.437
0.432
0.425
0.405
0.438
.0133 mbar
0.4555
0.429
0.441
0.453
0.462
0.415
0.488
0.48
0.47
0.433
0.492
.0266 mbar
0.5226
0.477
0.497
0.521
0.536
0.451
0.581
0.566
0.549
0.484
0.584
.0660 mbar
0.6819
0.595
0.637
0.679
0.705
0.544
0.778
0.764
0.727
0.608
0.796
0.13 mbar
0.878
0.745
0.814
0.868
0.9
0.668
1.009
0.99
0.944
0.768
1.053
0.26 mbar
1.1552
0.962
1.068
1.141
1.179
0.847
1.315
1.291
1.265
1.002
1.392
0.66 mbar
1.6833
1.386
1.589
1.664
1.668
1.194
1.826
1.805
1.914
1.469
2.014
1.33 mbar
2.2168
1.818
2.164
2.195
2.172
1.536
2.257
2.247
2.603
1.976
2.632
2.66 mbar
2.8418
2.333
2.939
2.814
2.695
1.921
2.647
2.666
3.508
2.631
3.313
6.66 mbar
3.6753
3.028
4.387
3.672
3.316
2.429
3.029
3.09
5.059
3.715
13.3 mbar
4.2056
3.48
5.774
4.225
3.67
2.734
3.204
3.33
6.361
4.605
4.699
26.6 mbar
4.5766
3.801
7.314
4.62
3.903
2.966
3.308
3.414
5.406
5.172
66.6 mbar
4.8464
4.037
4.916
4.071
3.075
3.43
3.509
6.159
5.583
133 mbar
4.9449
4.122
5.026
4.154
3.134
3.618
3.66
6.483
5.72
266 mbar
5.019
4.192
5.106
4.336
3.269
3.827
3.883
6.661
5.86
400 mbar
5.1111
4.283
5.2
4.502
3.384
3.938
4.005
6.726
533 mbar
5.2236
4.386
5.315
4.621
3.466
4.016
4.088
6.767
6.103
666 mbar
5.3294
4.477
5.422
4.708
3.526
4.076
4.151
6.803
800 mbar
5.4194
4.55
5.515
4.775
3.573
4.124
4.203
6.843
6.342
933 mbar
5.4949
4.611
5.592
4.83
3.613
4.166
4.247
6.89
1010 mbar
5.534
4.643
5.633
4.86
3.632
4.19
4.271
6.92
1060 mbar
5.5581
4.663
5.658
4.877
3.645
4.203
4.286
6.942
6.519
1190 mbar
5.6141
4.706
5.713
4.919
3.674
4.237
4.321
7
1330 mbar
5.6593
4.745
5.762
4.955
3.69
4.270
4.354
7.056
6.642
Values listed under each gas type are in volts.
Note: By design, these values are identical to the outputs from Brooks Automation, Inc. / Granville-Phillips
®
Convectron
®
gauges, Mini-Convectron
®
modules and controllers so that equivalent units can be interchanged
without affecting your process system or software.
An analog output of less than 0.01 volts to near 0 volt indicates a damaged or faulty sensor.