34FM-100 Power Requirements and Cords
Product Support:
34FM-100 Power Requirements and Cords
Power Requirements
lists the power requirements for the preferred 3-phase frame
configuration.
lists the power requirements for the optional single phase frame
configuration.
shows the Wye 3-phase configuration and
shows the Delta 3-phase configuration.
The 3-phase configuration uses a 4-wire connection as follows:
•
(L1) Line 1/Phase A
•
(L2) Line 2/Phase B
Table 5. 34FM-100 power requirements - 3-phase option
Parameter
Specification
Maximum Power (VA)
2400
3-Phase Voltage Wye or Delta (Vac) (±10%)
1
208 to 240
Recommended circuit breaker (required between
the facility power source and the load frame to
accommodate high in-rush currents at switch on).
3 pole, 10 A to 20 A
Frequency - Hz
47 to 63
1. Voltages are measured phase-to-phase. They are NOT measured between phase and ground or
neutral.
Table 6. 34FM-100 power requirements - single phase option
Parameter
Specification
Maximum Power (VA)
2400
Single Phase Voltage (Vac) (±10%)
208 to 240
Recommended circuit breaker (required between
the facility power source and the load frame to
accommodate high in-rush currents at switch on).
1 or 2 pole, 15 A to 20 A
Frequency - Hz
47 to 63
Содержание 34FM-100
Страница 1: ...Model 34FM 100 Dual Column Floor Model Preinstallation Manual M10 17541 EN Revision A...
Страница 8: ...Preliminary Pages 8 M10 17541 EN...
Страница 12: ...Preliminary Pages 12 M10 17541 EN...
Страница 50: ...Chapter Specifications 50 M10 17541 EN...
Страница 58: ...Chapter Risk reduction and safe use 58 M10 17541 EN...
Страница 72: ...Chapter Lifting and handling 72 M10 17541 EN...
Страница 74: ...74 M10 17541 EN...
Страница 75: ......
Страница 76: ...www instron com...