![Instron 34FM-100 Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/instron/34fm-100/34fm-100_preinstallation-manual_2070133028.webp)
Chapter: Requirements
28
M10-17541-EN
Ceiling clearance
Ensure that there is adequate ceiling clearance so that the load frame can be easily
placed in its testing location. Take into consideration how you will be transporting the
frame (by forklift or crane, for instance) and ensure that the ceiling height can
accommodate your mode of transportation.
Floor model frames are packed and shipped in a horizontal position. Make sure the
ceiling height is sufficient to allow righting the frame from the horizontal to a vertical
position (refer to
).
Environmental conditions
Ensure that the testing site meets the specifications detailed in
Accessibility
The testing system must be accessible for routine maintenance without having to move
the load frame. If the frame is moved, it may need to be re-verified (refer to
Site requirements for system accessories
Some accessories require additional site requirements. If you are not sure if the
accessories you purchased require any of the following site preparations, contact
Instron
®
Service or your local Instron
®
office for advice.
Dry compressed air
A dry, compressed air supply regulated to a maximum pressure of 8.3 bar (120 psi) is
required for the integrated air kit. The regulated supply pressure must be greater than
the test pressure value set in Bluehill
®
software.
The customer is responsible for adapting the pneumatic connection of the integrated air
kit to their supply. Instron
®
provides a 0.25 inch (6.3 mm) outer diameter, 0.125 inch
(3.2 mm) inner diameter tube, which can be connected using an appropriately sized
hose barb, compression or push-to-connect fitting. A 0.25 inch (6.3 mm) NPT male to
push-to-connect fitting is also supplied.
Содержание 34FM-100
Страница 1: ...Model 34FM 100 Dual Column Floor Model Preinstallation Manual M10 17541 EN Revision A...
Страница 8: ...Preliminary Pages 8 M10 17541 EN...
Страница 12: ...Preliminary Pages 12 M10 17541 EN...
Страница 50: ...Chapter Specifications 50 M10 17541 EN...
Страница 58: ...Chapter Risk reduction and safe use 58 M10 17541 EN...
Страница 72: ...Chapter Lifting and handling 72 M10 17541 EN...
Страница 74: ...74 M10 17541 EN...
Страница 75: ......
Страница 76: ...www instron com...