29
Environmental specifications
Product Support:
Water supply
A water supply may be required for some specimen grips and some temperature control
systems.
Liquid nitrogen and carbon dioxide
A liquid nitrogen (LN2) or carbon dioxide (CO2) supply may be required for
environmental chambers with cooling options.
For carbon dioxide, tank pressure must be approximately 725 psi. It must be in a liquid
state when it reaches the valve of the environmental chamber.
For liquid nitrogen, tank pressure must be between 5 and 10 psi, depending upon the
regulator and relief valves that are fitted to the dewar flask. Do not exceed 20 psi.
Environmental specifications
General
Conformity with EU directives
Instron
®
declares under our sole responsibility that 34FM testing systems are in
conformity with all relevant provisions of the following regulations:
•
Machinery Directive 2006/42/EC
•
EMC Directive 2014/30/EU
Table 4. Environmental specifications
Parameter
Specifications
Operating
Temperature
+5°C to +40°C (+41°F to +104°F)
Storage
Temperature
-25°C to +55°C (-13°F to +131°F)
Humidity
10% to 90% (non-condensing)
Ingress Protection (IP)
Rating
IP 2X. Protective measures may be required if excessive dust, corrosive fumes,
electromagnetic fields or hazardous conditions are encountered.
Содержание 34FM-100
Страница 1: ...Model 34FM 100 Dual Column Floor Model Preinstallation Manual M10 17541 EN Revision A...
Страница 8: ...Preliminary Pages 8 M10 17541 EN...
Страница 12: ...Preliminary Pages 12 M10 17541 EN...
Страница 50: ...Chapter Specifications 50 M10 17541 EN...
Страница 58: ...Chapter Risk reduction and safe use 58 M10 17541 EN...
Страница 72: ...Chapter Lifting and handling 72 M10 17541 EN...
Страница 74: ...74 M10 17541 EN...
Страница 75: ......
Страница 76: ...www instron com...