![Instron 2716-010 Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/instron/2716-010/2716-010_reference-manual_2070111034.webp)
Chapter: Installation
34
M10-14052-EN
Equipment
You need the following items:
•
a specimen capable of withstanding a tensile load equal to the current load string
preload
•
spanner wrench (1/4 in.)
Checklist
Check for the following conditions before you preload the grips:
•
The jaw faces are installed.
•
The side plates are secure. If you are using a 50 kN wedge grip, ensure that either
the spacers or side plates are fitted on the grip and are secure.
•
If using a 50 kN grip, check the thickness of your specimen and ensure that the
appropriate spacers or side plates are installed. Refer to
for additional information.
•
The Load channel indicates a zero load.
•
The electronic load limits are set to protect the weakest load string component.
Procedure
1. Install a specimen. Refer to section
2. Set the control electronics for a tensile load equal to the amount of the current load
string preload.
3. Using the spanner wrench, turn each locknut away from the mounting device (load
cell, baseplate clevis, or actuator).
4. Repeat
through
for the other grip.
5. Refer to section
to remove the specimen.
Содержание 2716-010
Страница 8: ...Preliminary Pages 8 M10 14052 EN...
Страница 14: ...Chapter Introduction 14 M10 14052 EN Description Figure 2 Grip Components 2716 015...
Страница 18: ...Chapter Introduction 18 M10 14052 EN...
Страница 24: ...Chapter Specifications 24 M10 14052 EN...
Страница 43: ...43 Parts Lists Product Support www instron com Parts Lists Figure 11 Illustrated Parts...
Страница 51: ......
Страница 52: ...Product Support www instron com...