Installation of the 2603-084 or 2603-086
Product Support:
Installation of the 2603-084 or 2603-086
Installation of the 2603-084 or 2603-086 onto 3300, 4400, 5500 and 5900 Series
Single Column Machines
1. Remove the screw located at the bottom of the front left-hand T-slot of the testing
machine.
2. Position the metal T-nut at the bottom of this T-slot with the threaded hole at the top of
the T-nut.
3. Replace the screw in the testing machine.
4. Press the M6 nut into the back of the plastic T-nut and position this at the top of the same
T-slot.
5. Remove the keep plate from the extensometer base plate. Keep the extensometer upright
once the counter-weights have been released. Re-fit the keep plate to the base plate in
the operating position.
6. Secure the extensometer head plate to the upper T-nut using the longer thumbscrew and
M6 nylon washer.
7. Secure the extensometer base plate to the lower T-nut using the shorter thumbscrew and
M6 nylon washer.
8. Remove the thumb nuts, washers and bolts from the arms. Place the arms onto the
carriages as shown. Replace the bolts, washers and thumb nuts.
9. The arms must be positioned on the carriages such that when the specimen is clamped,
the carriage support cables hang vertically (parallel to the guide rod).
10. Plug the extensometer connector plug into the STRAIN socket of the testing machine.